Validation of the STERIS Amsco 630LS steam sterilizer autoclave for inactivation of category a medical waste from patients with high-consequence infectious diseases

Published online by Cambridge University Press:  07 September 2023

Jessica A. Tarabay*
Affiliation:
Emory Healthcare, Atlanta, Georgia
Jill S. Morgan
Affiliation:
Emory Healthcare, Atlanta, Georgia
Colleen S. Kraft
Affiliation:
Division of Infectious Diseases, Emory University School of Medicine, Atlanta, GeorgiaDepartment of Pathology and Laboratory Medicine Emory University School of Medicine, Atlanta, Georgia
Mary Elizabeth Sexton
Affiliation:
Division of Infectious Diseases, Emory University School of Medicine, Atlanta, Georgia
*
Corresponding author: Jessica A. Tarabay; Email: Jessica.tarabay@emoryhealthcare.org
Rights & Permissions [Opens in a new window]

Abstract

Hospitals caring for patients with high-consequence pathogens may need to safely manage large volumes of category A waste. Using biological indicators to assess for successful sterilization, autoclave cycle parameters that would inactivate 4 categories of waste were identified and validated utilizing a STERIS Amsco 630LS Steam Sterilizer.

Information

Type
Concise Communication
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2023. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Table 1.Optimized Cycle Parameters by Waste Type

Figure 1

Figure 1.(A) Linen cycle temperature comparison. (B) Solidified liquid waste temperature comparison.