Quantification of intra-fraction motion in breast radiotherapy using cine EPID images

Published online by Cambridge University Press:  27 March 2024

Behnaz Fouladinezhad
Affiliation:
Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran
Mehdi Momennezhad
Affiliation:
Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran Nuclear Medicine Research Center, Mashhad University of Medical Sciences, Mashhad, Iran
Parisa Farjami
Affiliation:
Department of Biology, Islamic Azad University, Mashhad, Iran
Shahrokh Naseri*
Affiliation:
Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran
*
Corresponding author: Shahrokh Naseri; Email: naserish@mums.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

Background:

The aim of this study was to develop a method for accurately measuring the intra-fraction motion in cine electronic portal imaging device (EPID) images of tangential breast irradiation.

Patients and methods:

The cine EPID images were acquired for 18 breast cancer patients during medial tangential breast radiotherapy. The skin surface and the chest wall were segmented separately in each EPID image using an automated MATLAB algorithm to obtain the magnitude of intra-fraction motion.

Results:

The patients had an average age of 55 years (range 43–69 years), with the highest observed displacement at 3·50 mm. The mean intra-fraction motion due to respiration for the skin surface and chest wall was 1·04 ± 0·24 mm and 1·10 ± 0·27 mm, respectively. The amount of intra-fraction motion showed no significant association with either the patient’s age or the side of the breast receiving the treatment.

Conclusion:

In this study, by utilising cine EPID technology and the automated MATLAB algorithm, the intra-fraction motion during breast radiotherapy treatment was accurately measured and analysed. Results indicate minimal skin surface and chest wall motion (approximately 1 mm), confirming negligible intra-fraction motions during conformal radiotherapy in breast cancer patients.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Table 1. Patients’ clinical features and mean intra-fraction motion estimation

Figure 1

Figure 1. A sample of edge detection step. (a) Main EPID image obtained from a patient; (b) Result of the skin surface detection; (c) Result of the chest wall detection.

Figure 2

Figure 2. A sample of motion vector estimation step. (a) Result of the skin surface motion vector; (b) Result of the chest wall motion vector.

Figure 3

Table 2. Mean error of four motion estimation algorithms

Figure 4

Table 3. Mean and standard deviation of intra-fraction displacement of the skin surface and the chest wall, resulting from four methods of block-matching algorithms in the tangential radiation field (mm)

Figure 5

Figure 3. The linear relationship between the displacement of the skin surface and the chest wall.