Performance Evaluation of Spanish Biological Dosimetry Laboratories Through Interlaboratory Comparisons

Published online by Cambridge University Press:  13 October 2026

Anna Francès-Abellán
Affiliation:
Animal Biology, Plant Biology and Ecology, Universitat Autònoma de Barcelona , Catalonia, Spain
Alegria Montoro
Affiliation:
Radiological Protection Service, La Fe University and Polytechnic Hospital , Spain
Julien Alcaide
Affiliation:
Department of Radiotherapy, La Fe University and Polytechnic Hospital , Spain
Miguel Alcaraz
Affiliation:
Radiology and Physical Medicine Department, University of Murcia , Spain
Gemma Armengol
Affiliation:
Animal Biology, Plant Biology and Ecology, Universitat Autònoma de Barcelona , Catalonia, Spain
Inmaculada Domínguez
Affiliation:
Department of Cell Biology, University of Seville , Spain
José Luis Fernández García
Affiliation:
Molecular Genetics and Radiobiology Laboratory, Centro Oncológico de Galicia , Spain
Celia Fernández Martín
Affiliation:
Servicio de Oncología Radioterápica; Laboratorio de dosimetría biológica, Hospital General Universitario Gregorio Marañón , Spain
José Antonio García Gamuz
Affiliation:
Radiology and Physical Medicine Department, University of Murcia , Spain
Isidoro López Baltar
Affiliation:
Molecular Genetics and Radiobiology Laboratory, Centro Oncológico de Galicia , Spain
Andrei Cristian Marin
Affiliation:
Department of Radiotherapy, La Fe University and Polytechnic Hospital , Spain
Ana María Mercado Díaz
Affiliation:
Radiology and Physical Medicine Department, University of Murcia , Spain
Mercedes Moreno Domene
Affiliation:
Servicio de Oncología Radioterápica; Laboratorio de dosimetría biológica, Hospital General Universitario Gregorio Marañón , Spain
Amparo Olivares
Affiliation:
Radiology and Physical Medicine Department, University of Murcia , Spain
Nuria Pastor
Affiliation:
Department of Cell Biology, University of Seville , Spain
Jose Perez-Calatayud
Affiliation:
Department of Radiotherapy, La Fe University and Polytechnic Hospital , Spain
Maria Jesus Prieto
Affiliation:
Servicio de Oncología Radioterápica; Laboratorio de dosimetría biológica, Hospital General Universitario Gregorio Marañón , Spain
Miguel José Ruiz Gómez
Affiliation:
Department of Radiology and Physical Medicine, Universidad de Málaga , Spain
Albert Sánchez-Paños
Affiliation:
Department of Radiotherapy, La Fe University and Polytechnic Hospital , Spain
Juan Ignacio Villaescusa
Affiliation:
Radiological Protection Service, La Fe University and Polytechnic Hospital , Spain
Joan Francesc Barquinero*
Affiliation:
Animal Biology, Plant Biology and Ecology, Universitat Autònoma de Barcelona , Catalonia, Spain
*
Corresponding author: Joan Francesc Barquinero; Email: francesc.barquinero@uab.cat
Rights & Permissions [Opens in a new window]

Abstract

Objective

This study reports two interlaboratory comparisons (ILCs) dedicated to harmonizing biological dosimetry laboratories in Spain.

Methods

Two established techniques were applied among six laboratories: the dicentric chromosome assay (DCA) and the cytokinesis-block micronucleus (CBMN) assay. In the first ILC, blood samples were irradiated ex vivo at 2.9 Gy and 1.4 Gy. In the second ILC, a 2.63 Gy exposure was delivered generating a whole-body sample and a partial-body sample.

Results

In triage conditions using DCA, scoring 20 cells was sufficient to accurately classify doses above 2 Gy (i.e., high exposure) for rapid clinical prioritization, and 50 cells generally provided reliable estimates inside the $ \pm $0.5 Gy acceptance limit. The laboratories which could evaluate 500 cells obtained results where the 95% confidence intervals included the delivered dose. For partial-body exposures, scoring more than 50 cells was required at the tested dose (2.63 Gy) to ensure reliable DCA estimates. The CBMN assay successfully assessed whole-body exposure in the second exercise but could not identify partial-body exposure.

Conclusions

These ILCs demonstrate a general consistency among participant laboratories for whole-body dose assessment using both DCA and CBMN. However, further optimization is required for reliable application in partial-body triage scenarios.

Information

Type
Original Research
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - SA
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike licence (http://creativecommons.org/licenses/by-nc-sa/4.0), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the same Creative Commons licence is used to distribute the re-used or adapted article and the original article is properly cited. The written permission of Cambridge University Press or the rights holder(s) must be obtained prior to any commercial use.
Copyright
© The Author(s), 2026. Published by Cambridge University Press on behalf of Society for Disaster Medicine and Public Health, Inc
Figure 0

Figure 1. Map showing the geographical locations of the participant laboratories.Figure 1. long description.

Figure 1

Table 1. Cell culture specifications by laboratoryTable 1. long description.

Figure 2

Table 2. Calibration curves by laboratoryTable 2. long description.

Figure 3

Figure 2. Plots of the different dose estimations of the first ILC with 20, 50, and 500 analyzed cells by laboratory. Dose ranges are defined as: Low exposures (<1 Gy) in green, medium exposures (1-2 Gy) in yellow, and high exposures (>2 Gy) in red. a-c: Sample A simulates a whole-body irradiation of 2.9 Gy. d-f: Sample B simulates a whole-body irradiation of 1.4 Gy.Figure 2. long description.

Figure 4

Figure 3. Plots of the different dose estimations of the second ILC with 50 and 500 cell counts by laboratory. Dose ranges are defined as: Low exposures (<1 Gy) in green, medium exposures (1-2 Gy) in yellow, and high exposures (>2 Gy) in red. a-b: Sample A simulates a partial-body irradiation of 2.63 Gy. The whole-body equivalent was calculated for comparison. c-d: Sample B simulates a whole-body irradiation of 2.63 Gy.Figure 3. long description.

Figure 5

Figure 4. Plots of the Z-score value for samples A and B of the first ILC (a-b) and the second ILC (c-d) by laboratory. The color legend indicates the performance.Figure 4. long description.