Impact of modified techniques of transcatheter closure in large atrial septal defects (⩾30 mm) with anatomic complexities

Published online by Cambridge University Press:  23 July 2018

Ajith Ananthakrishna Pillai*
Affiliation:
Department of Cardiology, Jawaharlal Institute of Post-Graduate Medical Education and Research, Puducherry, India
Amarnath Upadhyay
Affiliation:
Department of Cardiology, Jawaharlal Institute of Post-Graduate Medical Education and Research, Puducherry, India
Saranya Gousy
Affiliation:
Department of Cardiology, Jawaharlal Institute of Post-Graduate Medical Education and Research, Puducherry, India
Amit Handa
Affiliation:
Department of Cardiology, Jawaharlal Institute of Post-Graduate Medical Education and Research, Puducherry, India
*
Author for correspondence: A. A. Pillai, Additional Professor, Department of Cardiology, Jawaharlal Institute of Post-Graduate Medical Education and Research, Puducherry 605006, India. Tel: +91 9488828946; Fax: +91 4137199; E-mail: ajithanantha@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Objective

The outcome of transcatheter closure in ostium secundum atrial septal defects is determined by the morphology of the defects. Modified techniques such as balloon assistance, pulmonary vein deployment, left atrial roof technique, and so on are used for circumventing the anatomic complexities and increasing the success rates.

Methods

We planned a prospective study looking at the outcomes of transcatheter closure in secundum atrial septal defects with modified techniques in different anatomic complexities identified in transoesophageal echocardiography and their association with outcome of transcatheter closure.

Results

Transcatheter closure was successful in 295 out of 346 (82%) patients with modified techniques. Balloon-assisted technique offered a success rate of 87%. The mean defect size was 34.7±2.78 mm (95% confidence interval (CI) 30.67–43.1 mm) with success and 40.16±4.5 mm (95% CI 32.16–44.7) with failure (p = 0.02). The mean total septal length was 38.11±0.63 (95% CI 35.21–40.56 mm) with success and 42.54±0.34 (95% CI 38.79–43.21 mm) with failure. The defect to septal ratios were 0.82 and 0.94 in success and failure groups, respectively (p=0.02). However, the absence of a retro-aortic margin, septal aneurysm, and multiple defects did not affect the success rate. Deficient inferior vena caval margin, deficient posterior margin, and size⩾40 mm had a high risk of failure with transcatheter closure. The odds ratio for procedural failure was 25.3 (4.3–143.8) in patients with malaligned septum, 8.3(1.4–48.5) with deficient inferior vena caval margin, and 4.1(2.5–19) for size⩾40 mm.

Conclusions

The modified techniques for device deployment offer substantial chances of success in transcatheter closure of secundum atrial septal defects with anatomical complexity (82%). Variants such as defect size of⩾40 mm and deficient inferior and posterior margins have high failure rates with a modified technique.

Information

Type
Original Article
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 Serial transoesophageal echocardiography images of balloon-assisted device closure of a large 38.5-mm atrial septal defect with deficient posterior margin (a to f). (a) Very large atrial septal defects measuring 38.5 mm with deficient posterior margin. (b) Sizing balloon occluding the defect. (c) Device in hour-glass shape with either discs in respective atria and waist compressed by inflated balloon. (d) Balloon is deflated and waist expands. (e) Device in good position snugly holding on after removal of the balloon. (f) Device after deployment.

Figure 1

Figure 2 Serial fluoroscopy images of transcatheter closure of a 32-mm atrial septal defect with a 36-mm device. Retro-aortic and inferior vena caval rims were deficient. The device embolised a few minutes later into the pulmonary artery. It was successfully retrieved and the same device was deployed with balloon-assisted technique with a snare holding the device before release. The procedure was successful with balloon-assisted technique.

Figure 2

Table 1 Baseline characteristics.

Figure 3

Table 2 Modified techniques for device delivery.

Figure 4

Figure 3 Pulmonary vein (PV) deployment technique in a patient with malaligned septum (a to d). (a) Transthoracic echo showing a posterior malalignment of the septum. The dotted lines show the malalignment. (b) The left atrial disc is deployed in the PV and the device is stretched across the defect. (c) Image showing stretched device with left atrial disc anchored to the left upper pulmonary vein. (d) The device in position after retraction from PV.

Figure 5

Figure 4 The left atrial roof method. (a) The left atrial disc pushed against the atrial wall and the device stretched across the defect to open out the right atrial disc as well. (b) The device is stretched across the atrial septal defect and the right atrial (RA) disc and pushed out. (c) Once RA disc is in position, a gentle pull on the cable causes the left atrial disc to fall back. (d) Device in position after release.

Figure 6

Figure 5 Dilator-assisted method in cases of small left atrium. (a) Dilator with wire in position in the left atrium (LA) with the LA disc opened. (b) Device deployed in alignment with the support of a dilator.

Figure 7

Figure 6 (a) Transthoracic echo image with extreme posterior septal malalignment (“S”-shaped septum) with multiple atrial septal defects (arrows). (b) Anterior and posterior colour flow indicating multiple shunts. (c) Fluoroscopic image showing catheters crossing multiple defects under transoesophageal guidance. (d) Balloon-assisted delivery of the first device; the balloon is being inflated in the anterior defect to counter the malalignment and to prevent device prolapse across the posterior defect. (e) Subsequent successful closure of defects with two devices. The devices are oriented in different directions, indicating the extreme malalignment of the septum.

Figure 8

Table 3 Success rates in various techniques in different unfavourable morphologies (successful/attempted).

Figure 9

Table 4 Prediction model for intervention failure.

Figure 10

Figure 7 Late erosion of the device into the aorta in a patient with malaligned defect. (a) Four-chamber view showing the malaligned septum. (b) Parasternal long-axis view immediately after implantation of a 36-mm device. (c) Late erosion into the aorta. (d) Colour flow showing the eroded area with aortic regurgitation.

Figure 11

Figure 8 Pictorial representation of various techniques. (a) Serial diagrams showing the steps of balloon assistance method. (b) Pulmonary vein method. (c) Left atrial (LA) roof technique. (d) Dilator-assisted technique. LV=left ventricle; RA=right atrial; RV=right ventricle; TV=tricuspid valve.