An infrared interactive patient position guidance and acquisition control system for use during radiotherapy treatment

Published online by Cambridge University Press:  06 April 2017

Hiraku Fuse*
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Inashiki-gun, Ibaraki, Japan
Kenji Komatsu
Affiliation:
Department of Radiological Technology, Hitachi General Hospital, Hitachi, Ibaraki, Japan
Hiroyuki Arakawa
Affiliation:
Department of Radiological Technology, Seirei Sakura Citizen Hospital, Sakura, Chiba, Japan
Takeji Sakae
Affiliation:
Faculty of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan
Fujisaki Tatsuya
Affiliation:
Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Inashiki-gun, Ibaraki, Japan
*
Correspondence to: Hiraku Fuse, Department of Radiological Sciences, Ibaraki Prefectural University of Health Science, 4669-2 Ami-machi, Inashiki-gun, Ibaraki 300-0331, Japan. Tel: +81 298 40 2218. Fax: +81 298 40 2318. E-mail: fuseh@ipu.ac.jp
Rights & Permissions [Opens in a new window]

Abstract

Background

The control of patient position, posture and respiratory movements during radiotherapy is important for effective and specific treatment of malignancy. We have developed an infrared (IR) interactive patient position guidance and acquisition control system for clinical use, comprising IR cameras, IR markers and dedicated software.

Materials and methods

We evaluated the system with ten healthy volunteers and ten experienced operators. IR markers were placed on the body surface. Their positions were calculated using vectors of three translational and three rotational parameters, and the intrafractional error for each marker was acquired with and without respiratory motion. The inclusion of multiple positioning markers allowed for real-time visualisation of the patient posture, with feedback on misalignment and required postural adjustments.

Results

The positioning time was 73 seconds (with a minimum period of 39 seconds), which was significantly shorter than for conventional line alignment. A comparison of positioning reproducibility between conventional line alignment and this system was <3·5 mm and was not patient dependent or operator dependent. An intrafractional error of displacement of up to 10·0 mm was found in the right iliac crest.

Conclusions

This IR interactive system was shown to be high utility and suitable for monitoring patient position, posture and respiratory movements during radiotherapy.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1 The left side of figure shows the placement of infrared camera. Black squares indicate the treatment couch, which is also the region of acquire for the displacement of each marker from its reference position. (a), (b) and (c) are positioning process. The function of each arrow is to change the size by quantity of displacement with the reference position.

Figure 1

Figure 2 The mean infrared (IR) marker attachment time. Notes: The plots are result of positioning by ten operators. Solid line indicates the mean IR marker attachment time.

Figure 2

Figure 3 The mean infrared (IR) marker attachment accuracy. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean IR marker attachment accuracy.

Figure 3

Figure 4 A comparison between the two datasets to determine whether positioning differed in the craniocervical system between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment accuracy. Abbreviation: IR, infrared.

Figure 4

Figure 5 A comparison between the two datasets to determine whether positioning differed in the torso between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment accuracy. Abbreviation: IR, infrared.

Figure 5

Figure 6 A comparison between the two datasets to determine whether positioning differed in the pelvice between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment accuracy. Abbreviation: IR, infrared.

Figure 6

Figure 7 A comparison between the two datasets to determine whether positioning time differed in the craniocervical system between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment time. Abbreviation: IR, infrared.

Figure 7

Figure 8 A comparison between the two datasets to determine whether positioning time differed in the torso between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment time. Abbreviation: IR, infrared.

Figure 8

Figure 9 A comparison between the two datasets to determine whether positioning time differed in the pelvice between experimental laser centring procedures and the patient positioning guidance system. Notes: The plots are results of positioning by ten operators. Solid line indicates the mean patient alignment time. Abbreviation: IR, infrared.