In vivo experiences with magnetic resonance imaging scans in Vibrant Soundbridge type 503 implantees

Published online by Cambridge University Press:  23 April 2018

I Todt*
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
P Mittmann
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
A Ernst
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
S Mutze
Affiliation:
Department of Radiology, Unfallkrankenhaus Berlin, Germany
G Rademacher
Affiliation:
Department of Radiology, Unfallkrankenhaus Berlin, Germany
*
Author for correspondence: Dr Ingo Todt, Department of Otolaryngology, Unfallkrankenhaus Berlin, Warener Str. 7, D-12683 Berlin, Germany Fax: +49 30 5681 4303 E-mail: todt@gmx.net
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To observe the effects of magnetic resonance imaging scans in Vibrant Soundbridge 503 implantees at 1.5T in vivo.

Methods:

In a prospective case study of five Vibrant Soundbridge 503 implantees, 1.5T magnetic resonance imaging scans were performed with and without a headband. The degree of pain was evaluated using a visual analogue scale. Scan-related pure tone audiogram and audio processor fitting changes were assessed.

Results:

In all patients, magnetic resonance imaging scans were performed without any degree of pain or change in pure tone audiogram or audio processor fitting, even without a headband.

Conclusion:

In this series, 1.5T magnetic resonance imaging scans were performed with the Vibrant Soundbridge 503 without complications. Limitations persist in terms of magnetic artefacts.

Information

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2018 
Figure 0

Table I Individual attachments of Vibrant Soundbridge

Figure 1

Table II Mean hearing thresholds pre- and post-scanning

Figure 2

Fig. 1 Magnetic resonance imaging artefact of a Vibrant Soundbridge 503 system in the (a) axial and (b) sagittal planes.