Novel Laser Ablation Sampling Device for the Rapid Radiocarbon Analysis of Carbonate Samples by Accelerator Mass Spectrometry

Published online by Cambridge University Press:  17 February 2016

C Welte
Affiliation:
Laboratory of Inorganic Chemistry, D-CHAB, ETHZ, Vladimir-Prelog Weg 1, 8093 Zurich, Switzerland. Laboratory of Ion Beam Physics, ETHZ, Otto-Stern Weg 5, HPK, 8093 Zurich, Switzerland.
L Wacker*
Affiliation:
Laboratory of Ion Beam Physics, ETHZ, Otto-Stern Weg 5, HPK, 8093 Zurich, Switzerland.
B Hattendorf*
Affiliation:
Laboratory of Inorganic Chemistry, D-CHAB, ETHZ, Vladimir-Prelog Weg 1, 8093 Zurich, Switzerland.
M Christl
Affiliation:
Laboratory of Ion Beam Physics, ETHZ, Otto-Stern Weg 5, HPK, 8093 Zurich, Switzerland.
J Koch
Affiliation:
Laboratory of Inorganic Chemistry, D-CHAB, ETHZ, Vladimir-Prelog Weg 1, 8093 Zurich, Switzerland.
H-A Synal
Affiliation:
Laboratory of Ion Beam Physics, ETHZ, Otto-Stern Weg 5, HPK, 8093 Zurich, Switzerland.
D Günther
Affiliation:
Laboratory of Inorganic Chemistry, D-CHAB, ETHZ, Vladimir-Prelog Weg 1, 8093 Zurich, Switzerland.
*
*Corresponding author. Email: wacker@phys.ethz.ch.
*Corresponding author. Email: wacker@phys.ethz.ch.
Rights & Permissions [Opens in a new window]

Abstract

Conventional radiocarbon analysis of carbonate records with accelerator mass spectrometry (AMS) is time consuming and the achievable spatial resolution is limited, because individual samples have to be taken and need to be converted to graphite for the measurement. A new laser ablation (LA) in situ sampling technique for rapid online 14C analyses of carbonate records by AMS is presented. By focusing a 193-nm ArF excimer laser on carbonate samples, carbon dioxide is generated and can directly be introduced into the gas ion source of an AMS. A dedicated LA cell for AMS was constructed in a way that combines rapid gas exchange with the capacity to carry sample specimen with maximum dimensions of 15×2.5×1.5 cm3. With the presented setup, negative carbon ion currents up to 20 µA were achieved. A low 14C background of 0.011±0.002 F14C was observed on 14C-free marble and different standard and reference materials could be well reproduced within errors. The novel technique allows scanning carbonate samples continuously over several cm per hour with achievable measurement precisions of less than 1% for modern samples. This approach allows acquiring highly spatially resolved 14C records at a far higher rate than with any currently available method.

Information

Type
Research Article
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Table 1 Typical laser settings during LA-AMS measurements.

Figure 1

Figure 1 Schematic overview of the LA-AMS setup. The LA cell consists of two parts: the ablation head (A) and the sample box (B). Black and dashed lines represent the vacuum system, gray lines the helium connections, and white lines the connection between the ablation head and the ion source. Legend: 1. flow adjustment, 2. pressure sensor, 3. window, 4. filter, 5. cap, 6. four-way valve, 7. LA-AMS capillary, 8. capillary in ion source,×shut-off valve.

Figure 2

Figure 2 Enlargement of the ablation head. The focused laser beam entering through the window at an angle of 45° is indicated. Furthermore, the ablation plume and emitted particles and resulting deposits on the cell wall are shown. The insert reduces the gap between the ablation head and the sample resulting in a higher flow resistance for gas from the ablation head to the sample box.

Figure 3

Table 2 Overview of the different materials used for LA-AMS and their 14C content.

Figure 4

Figure 3 The CO2 conversion efficiency of the Ex5 laser system as a function of the fluence. The solid line represents the mean value of the four measurements and the dashed lines indicate the standard error.

Figure 5

Figure 4 Ion current as a function of the C-flow rate. The error bars indicate the ion current range and, thus, represent the signal stability. The pressure in the cell was kept constant resulting in a gas flow rate of 1.5 mL/min. The maximum ion current is found at a C flow of ~4 µg/min.

Figure 6

Figure 5 Depending on the C-flow rate into the ion source, one of the following cases occurs: (a) Optimum conditions cause a quick rise in the 12C current, which is followed by a steady signal. (b) A low C flow results in a slow and steady signal rise. The current increase at 2.5 min is caused by an increase in the helium background pressure in the LA cell. For both cases, LA was used as a sample introduction technique.

Figure 7

Figure 6 Ion current dependence on the helium background pressure in the LA cell for two capillaries with different flow resistances. The maximum ion current is reached at different pressures for the two capillaries.

Figure 8

Table 3 Figures of merit of conventional gas measurements and of the LA-AMS system.

Figure 9

Figure 7 LA-blank measurements (black dots) during 3 days and blank gas passed directly through the LA cell on 2 days (squares). The open symbols represent the first 15 min of measurement on each day and were excluded from calculating the mean blank value and its standard deviation (gray area). The shaded area corresponds to the conventional gas blank level. Furthermore, the 12C current (triangles) corresponding to each blank measurement is depicted. Laser settings are equal to those in Table 1. The pressure in the LA cell was set as follows: 200–500 mbar (capillary 2) on day 1, 500–1100 mbar (capillary 1), and 300–900 mbar (capillary 2) on day 2, 550–1100 mbar (capillary 1) on day 3.

Figure 10

Figure 8 Washout behavior of the LA cell after lasing for 30 min, which is an exceptionally long lifetime fo the sputter target. The dashed line indicates the background level of the ion current, which normally reaches 0.1 µA after presputtering.

Figure 11

Figure 9 Study of the effect of cross-contamination during LA. Blanks measured directly after standards are elevated. The same can be observed for standards measured after blanks. The gray band represents the blank range normally yielded for LA-AMS measurements. For the standard materials, the standard deviation of the nominal value is shown (gray lines).

Figure 12

Figure 10 Measured F14C divided by the nominal value (referred to as “fraction nominal”) for three standard materials using LA-AMS. The open symbols represent data that are also shown in Figure 9. Solid vertical lines indicate different measurement days, while dashed lines represent the measurement of other samples in between. The solid horizontal line indicates the nominal value of the standards, dashed horizontal lines the standard deviation of the nominal value, and gray areas represent 1 standard deviation of the measured values.

Figure 13

Figure 11 Comparison of four scans on two sections of a bamboo coral with the mean F14C value across the growth axis from conventional graphite measurements (gray band). The arrows indicate where the scans were performed and the arrow heads point into the direction of scan.