Computed tomography scanning of implanted hypoglossal nerve stimulators and approach to device malfunction: case series

Published online by Cambridge University Press:  13 December 2021

M J Urban*
Affiliation:
Department of Otorhinolaryngology – Head and Neck Surgery, Rush University Medical Center, Chicago, USA
A Calder
Affiliation:
Drexel University College of Medicine, Philadelphia, USA
M Jhaveri
Affiliation:
Section of Neuroradiology, Department of Radiology, Rush University Medical Center, Chicago, USA
P LoSavio
Affiliation:
Department of Otorhinolaryngology – Head and Neck Surgery, Rush University Medical Center, Chicago, USA
*
Author for correspondence: Dr M J Urban, Department of Otorhinolaryngology – Head and Neck Surgery, Rush University Medical Center, 1611 W. Harrison St., Suite 550 Orthopedic Building, Chicago 60612, USA E-mail: matthew_j_urban@rush.edu
Rights & Permissions [Opens in a new window]

Abstract

Objective

Serious device-related complications for hypoglossal nerve stimulators are rare, but surgeons should implement a prompt and systematic approach to quickly troubleshoot a non-functioning device.

Method

Records were queried at a single academic tertiary referral centre between January 2019 and June 2021.

Results

The authors present four cases of non-functioning hypoglossal nerve stimulator devices: one case in which migration of the stimulation lead required a revision implantation, one in which the implantable pulse generator was found to be non-functional intra-operatively, one case of an intramuscular sensory lead tract causing pain and one case of implantable pulse generator failure that was probably triggered by implantable cardiac device discharge. In this study, computed tomography imaging was critical to the diagnosis for the first and third cases.

Conclusion

Given the limited complication reporting available for hypoglossal nerve stimulators, these cases highlight management and unique imaging findings. The authors present an algorithm to work-up non-functioning hypoglossal nerve stimulator devices.

Information

Type
Clinical Records
Copyright
Copyright © The Author(s), 2021. Published by Cambridge University Press on behalf of J.L.O. (1984) LIMITED
Figure 0

Fig. 1. Computed tomography (CT) scans of the neck for case 1. (a) Axial plane CT scan of the neck (arrowhead indicates wire). (b) Axial plane CT scan of the neck (arrow indicates cuff). (c) Coronal plane CT scan of the neck (arrowhead indicates wire). (d) Coronal plane CT scan of the neck (arrow indicates cuff). (e) Sagittal plane CT scan of the neck (arrowhead indicates wire; arrow indicates cuff). (f) Three-dimensional CT scan reconstruction demonstrating the cuff is clearly oriented more inferiorly and directed caudally in comparison to the expected course of the nerve. In comparison to Figure 3, the cuff is positioned more inferiorly and posteriorly, hanging below the anterior belly of the digastric muscle in a more caudal orientation (arrow indicates cuff). The anchor remains attached to the digastric muscle.

Figure 1

Fig. 2. Neck and chest computed tomography scan for a patient with pain because of lead tethering (case 2). (a) Axial computed tomography scan of the chest demonstrating the sensory lead (arrow) tunnelling through the pectoralis major muscle and coursing between the pectoralis major and minor muscle (b). The three-dimensional reconstruction (c) demonstrates the sensory lead's full (arrow) course through the pectoralis major muscle (hazy grey) to the intercostal space. (d) Axial computed tomography scan of the neck at the level of the thyroid cartilage with the stimulator wire (arrow) tunnelling within the sternocleidomastoid muscle at the site of the patient's pain.

Figure 2

Fig. 3. Computed tomography (CT) scans of the neck for a properly positioned hypoglossal nerve stimulator. (a) Axial plane CT scan of the neck (arrow indicates cuff). (b) Axial plane CT scan of the neck (arrowhead indicates wire). (c) Coronal plane CT scan of the neck (arrowhead indicates wire). (d) Coronal plane CT scan of the neck (arrow indicates cuff). (e) Sagittal plane CT scan of the neck (arrow indicates cuff; arrowhead indicates wire). (f) Three-dimensional reconstruction demonstrates the course of the lead and proper position of the stimulation cuff in relation to the mandible and hyoid bones. The radiopaque, three-pronged stimulation cuff (Inspire II) should be visualised in the inferior sublingual space, oriented anteriorly in the axial plane and mirror the expected course of the nerve (arrow indicates cuff). The wiring is tunnelled under the anterior belly of the digastric muscle, and the anchor device is secured to the muscle (arrowhead indicates wire).

Figure 3

Fig. 4. General algorithm for management of device malfunction at the author's institution. The ellipses indicate these are examples of common causes of device malfunction and are not intended to be all-encompassing. IPG = implantable pulse generator; CT = computed tomography; XII = hypoglossal nerve.