Quality assurance of linear accelerator: a comprehensive system using electronic portal imaging device

Published online by Cambridge University Press:  22 October 2018

P. Niyas*
Affiliation:
Department of Physics, Farook College, Kerala, India Department of Medical Physics, Baby Memorial Hospital, Kerala, India Department of Medical Physics, MVR Cancer Centre & Research Institute, Kerala, India
K. K. Abdullah
Affiliation:
Department of Physics, Farook College, Kerala, India
M. P. Noufal
Affiliation:
Department of Medical Physics, Baby Memorial Hospital, Kerala, India
R. Vysakh
Affiliation:
Department of Medical Physics, MVR Cancer Centre & Research Institute, Kerala, India
*
Author for correspondence: P. Niyas, Akayi Potta, Farook College, Calicut 673632, Kerala, India. Tel: +91 9847814131. E-mail: pniyas@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Aim

The Electronic Portal Imaging Device (EPID), primarily used for patient setup during radiotherapy sessions is also used for dosimetric measurements. In the present study, the feasibility of EPID in both machine and patient-specific quality assurance (QA) are investigated. We have developed a comprehensive software tool for effective utilisation of EPID in our institutional QA protocol.

Materials and methods

Portal Vision aS1000, amorphous silicon portal detector attached to Clinac iX—Linear Accelerator (LINAC) was used to measure daily profile and output constancy, various Multi-Leaf Collimator (MLC) checks and patient plan verification. Different QA plans were generated with the help of Eclipse Treatment Planning System (TPS) and MLC shaper software. The indigenously developed MATLAB programs were used for image analysis. Flatness, symmetry, output constancy, Field Width at Half Maximum (FWHM) and fluence comparison were studied from images obtained from TPS and EPID dosimetry.

Results

The 3 years institutional data of profile constancy and patient-specific QA measured using EPID were found within the acceptable limits. The daily output of photon beam correlated with the output obtained through solid phantom measurements. The Pearson correlation coefficients are 0.941 (p = 0.0001), 0.888 (p = 0.0188) and 0.917 (p = 0.0007) for the years of 2014, 2015 and 2016, respectively. The accuracy of MLC for shaping complex treatment fields was studied in terms of FWHM at different portions of various fields, showed good agreement between TPS-generated and EPID-measured MLC positions. The comparison of selected patient plans in EPID with an independent 2D array detector system showed statistically significant correlation between these two systems. Percentage difference between TPS computed and EPID measured fluence maps calculated for number of patients using MATLAB code also exhibited the validity of those plans for treatment.

Information

Type
Original Article
Copyright
© Cambridge University Press 2018 
Figure 0

Table 1 Customised QA protocol

Figure 1

Figure 1 Daily analysis of flatness and symmetry along X and Y profiles using MATLAB program.

Figure 2

Figure 2 Fluence maps of different MLC shapes created with the help of MLC shaper. First and third columns represent the fluence maps from TPS and second and fourth columns represent fluences measured by EPID.Abbreviations: EPID, Electronic Portal Imaging Device; MLC, Multi-Leaf Collimator; TPS, Treatment Planning System.

Figure 3

Figure 3 Comparison of TPS-computed and EPID-measured fluences of an MLC pattern (Z shape). FWHM is calculated and displayed.Abbreviations: EPID, Electronic Portal Imaging Device; FWHM, Field Width at Half Maximum; MLC, Multi-Leaf Collimator; TPS, Treatment Planning System.

Figure 4

Figure 4 Image comparison between TPS generated and EPID measured 2D fluences of an MLC shape and a patient-specific QA plan. Percentage difference and RMSE are also calculated by using MATLAB program.Abbreviations: EPID, Electronic Portal Imaging Device; MLC, Multi-Leaf Collimator; RMSE, root mean square error; QA, quality assurance; TPS, Treatment Planning System.

Figure 5

Figure 5 Three years institutional data of flatness and symmetry along X and Y profiles.

Figure 6

Figure 6 The comparison of 6 MV photon beam output constancy between daily EPID and weekly solid phantom measurement.Abbreviation: EPID, Electronic Portal Imaging Device.

Figure 7

Table 2 Detailed comparison report of field widths planned by MLC shaper, FWHM computed and measured by TPS and EPID for various MLC shapes at different locations

Figure 8

Table 3 Gamma results of fluence comparison between TPS computed vs. MatriXX 2-D measured for four different anatomic regions. % difference of pixels (TPS vs. EPID) computed by MATLAB code is also tabulated

Figure 9

Figure 7 Validation of image comparison MATLAB code by comparing three sets of images. From top to bottom row: TPS fluence of an MLC shape (L) and its own copy, TPS vs. EPID measured fluences of an MLC shape (L) and TPS vs. EPID measured fluences of two different MLC shapes (L and H).Abbreviations: EPID, Electronic Portal Imaging Device; MLC, Multi-Leaf Collimator; TPS, Treatment Planning System.

Figure 10

Table 4 Comparison results (% difference and RMSE) of different fluence shapes, calculated using MATLAB code. Large disagreement is observed in the comparison of unmatched images.