Dielectric contrast measurements on biological substances with resonant microwave near-field sensors

Published online by Cambridge University Press:  10 May 2013

Nora Haase*
Affiliation:
Institut für Hochfrequenztechnik, Technische Universität Hamburg-Harburg, 21073 Hamburg, Germany. Phone:  +49 (0)40-42878-2225
Arne F. Jacob
Affiliation:
Institut für Hochfrequenztechnik, Technische Universität Hamburg-Harburg, 21073 Hamburg, Germany. Phone:  +49 (0)40-42878-2225
*
Corresponding author: Nora Haase Email: nora.haase@tu-harburg.de
Rights & Permissions [Opens in a new window]

Abstract

Resonant substrate integrated near-field sensors are used for characterization of aqueous solutions at three different frequencies. In addition, Chinese hamster ovary (CHO) cells in a culture medium are characterized with the same sensors. Different concentrations as well as different vital states of cell suspensions are examined. The complex permittivity of the samples is evaluated using a linearized method based on perturbation theory. The permittivity differences between the measured cell suspensions are discussed. The resonant frequencies of the sensors are close to 3, 7, and 11 GHz, respectively.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2013 
Figure 0

Fig. 1. Sketch of a folded near-field sensor.

Figure 1

Table 1. Geometry parameters of the folded near-field sensors.

Figure 2

Fig. 2. Measurement setup for liquid samples using substrate integrated sensor.

Figure 3

Table 2. Resonator parameters of the sensors loaded with air.

Figure 4

Fig. 3. Measured resonator parameters for different aqueous solutions.

Figure 5

Table 3. Standard deviations of resonant frequencies and quality factors encountered in measurements with air, water, and saline solution samples with Sensor 1 for error discussion.

Figure 6

Table 4. Complex resonant frequencies and constants C used in the perturbation approach for calculation of complex permittivity.

Figure 7

Fig. 4. Measured and reference [15] permittivity values for different saline solutions.

Figure 8

Fig. 5. Reproducibility test with cell suspensions of different cultivation batches.

Figure 9

Fig. 6. Biological protocol for preparation of different cell concentrations after cultivation.

Figure 10

Fig. 7. Sketch of cell behavior in funnel, not to scale. ○1 Plexiglas funnel, sensing tip, seal to prevent leakage, seal to avoid evaporation, cells, evenly distributed, cells, sedimented.

Figure 11

Table 5. Estimated thickness of sedimented cells based on geometrical considerations.

Figure 12

Fig. 8. Complex permittivity of alive cells sedimenting on Sensors 1 and 2 in different concentrations.

Figure 13

Table 6. Permittivity and loss factor of sedimented cells and dielectric change during sedimentation in a cell suspension of c = 10 × 106 cells/ml.

Figure 14

Fig. 9. Biological protocol for preparation of different vital states of cells.

Figure 15

Fig. 10. Images of different vital states of CHO cell suspensions. The colouring indicates the vital state of the cells: light cells are alive, dark cells are dead. Cell fragments are not coloured, therefore not visible optically.

Figure 16

Fig. 11. Calculated complex permittivity of cell sedimentation using Sensors 1 and 3. The cell concentration is 10 × 106.

Figure 17

Table 7. Difference between permittivity of living and dead cells observed with the sensors.