Effect of Beam Transmission of Stable Isotopes on Online δ13C for SSAMS

Published online by Cambridge University Press:  23 July 2019

G V Ravi Prasad*
Affiliation:
Center for Applied Isotope Studies, University of Georgia, 120 Riverbend Rd., Athens, GA 30602, USA
Alexander Cherkinsky
Affiliation:
Center for Applied Isotope Studies, University of Georgia, 120 Riverbend Rd., Athens, GA 30602, USA
Randy Culp
Affiliation:
Center for Applied Isotope Studies, University of Georgia, 120 Riverbend Rd., Athens, GA 30602, USA
*
*Corresponding author. Email: gvrp@uga.edu.
Rights & Permissions [Opens in a new window]

Abstract

It is known that 12C beam transmission through the accelerator decreases at high beam currents. This effect depends on machine design and varies across different types of AMS instruments. For beam currents of about 100 μA, the effect is small on the 500 kV tandem CAMS unit, whereas beam saturation is observed for similar high beam currents on the 250 kV SSAMS unit. While this effect is very evident for high 12C beam currents, we have also observed that even the 13C beam is found to suffer modest transmission loss with beam current. As a result, the 13C/12C ratio does not remain constant with beam current. By correcting for the effects of 12C beam saturation and decreased 13C transmission, we have obtained online δ13C values that are more accurate and precise at moderately high beam currents for SSAMS.

Information

Type
Conference Paper
Copyright
© 2019 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Figure 1 (a) Post acceleration 12C+ vs. pre-acceleration 12C– for OXII and C6 standards. Data points correspond to individual runs, with errors smaller than the size of the points. Beam saturation of 12C+ may be seen above 80 μA. (b) In contrast the 13C+ vs. 12C– plot is linear. Only C6 data is plotted for clarity.

Figure 1

Figure 2 Difference between AMS and IRMS δ13C values plotted as histograms. (a) 13C+/12C– ratio results in better δ13C than 13C+/12C+ ratio. (b) 13C+/12C– ratio corrected for 13C+ transmission reduction results in a symmetric curve and closest agreement with IRMS values.

Figure 2

Figure 3 13C+ transmission decreases with beam current. The data points correspond to a single sample group with two OXII and two C6 samples. Data for <85 μA correspond to OXII. Data points for >85 μA correspond to C6 but scaled to match the 13C+/12C– ratio of OXII.

Figure 3

Figure 4 (a) pMC data of C6 calculated using δ13C = –10.8‰. The mean value 150.45 ± 0.62 of the data shown as a solid line agrees well with the nominal value of 150.6. Dashed lines show 1σ deviation from the mean value. (b) 14C+/13C+ ratio of C6 normalized to the OXII ratio does not change with the beam current. The data points correspond to individual runs, and the error bars of each point are smaller than the size of the points. The data is in exact agreement with the expected value of 1.13. Dashed lines show 1σ deviation from the mean value.