Is interventional closure the current treatment of choice for selected patients with deficient atrial septation?

Published online by Cambridge University Press:  03 February 2006

Per G. Bjørnstad
Affiliation:
Paediatric Cardiology, Rikshospitalet – The National Hospital, University of Oslo, Oslo, Norway
Rights & Permissions [Opens in a new window]

Abstract

Information

Type
Review
Copyright
© 2006 Cambridge University Press
Figure 0

(a) CardioSEAL device with two springs in the middle of each arm in addition to the spring at the origin of the arms from the centre of the device. The left-sided patch is placed at an angle of 45 degree to the right-sided Dacron patch. Placed like this, the arrangement is flat. The fluoroscopy frame (b) shows the device in the left anterior oblique view, still kept on the delivery system. The left-sided and right-sided arms are pushed apart by the thickness of the atrial septum making the whole device much “thicker”. The transoesophageal echo probe is readily visible at the top right.

Figure 1

The left-sided aspect of the circular Amplatzer device is shown in this photograph (a). The device is made of 72 thin nitinol wires and filled with three layers of Dacron fibres. The sutures keeping the fibres in place are well seen on this frame. The lighter circle in the middle indicates the central core being the part that plugs the defect. In the lateral view (b), the slimmer right-sided and the heavier left-sided disc are seen. They are supposed to adapt the device to the septum, so, that the central core – visible between the discs – fits into the atrial septal defect. The device is attached to the delivery cable with a microscrew.

Figure 2

The Helex device is a Gore-Tex non-centring double disc device. The Gore-Tex material covers one single nitinol wire, which supports the tissue and shapes the device. The lower frame shows its endothelialisation in the animal experiments, demonstrating full overgrowth of this very flat device.