Hostname: page-component-74656f9884-qr5l4 Total loading time: 0 Render date: 2026-11-04T02:31:42.803Z Has data issue: false hasContentIssue false

Clinical experience using Delta 4 phantom for pretreatment patient-specific quality assurance in modern radiotherapy

Published online by Cambridge University Press:  15 November 2018

R. P. Srivastava*
Affiliation:
Department of Radiation Oncology, Ghent University Hospital, Gent, Belgium Radiotherapy Association Meuse Picardie, Centre Hospitalier Mouscron, Mouscron, Belgium
C. De Wagter
Affiliation:
Department of Radiation Oncology, Ghent University Hospital, Gent, Belgium Department of Radiation Oncology and Experimental Cancer Research, Ghent University Hospital, Gent, Belgium
*
Author for correspondence: R. P. Srivastava, Department of Radiation Oncology, Ghent University Hospital, C. Heymanslaan 10, Radiotherapiepark, 9000 Gent, Belgium. Tel: +3293323030. E-mail: rajupsrivastava@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose

In advanced radiotherapy techniques such as intensity-modulated radiation therapy (IMRT), the quality assurance (QA) process is essential. The aim of the study was to assure the treatment planning dose delivered during delivery of complex treatment plans. The QA standard is to perform patient-specific comparisons between planned doses and doses measured in a phantom.

Materials and method

The Delta 4 phantom (Scandidos, Uppsala, Sweden) has been used in this study. This device consists of diode matrices in two orthogonal planes inserted in a cylindrical acrylic phantom. Each diode is sampled per beam pulse so that the dose distribution can be evaluated on segment-by-segment, beam-by-beam, or as a composite plan from a single set of measurements. Ninety-five simple and complex radiotherapy treatment plans for different pathologies, planned using a treatment planning system (TPS) were delivered to the QA device. The planned and measured dose distributions were then compared and analysed. The gamma index was determined for different pathologies.

Results

The evaluation was performed in terms of dose deviation, distance to agreement and gamma index passing rate. The measurements were in excellent agreement between with the calculated dose of the TPS and the QA device. Overall, good agreement was observed between measured and calculated doses in most cases with gamma values above 1 in >95% of measured points. Plan results for each test met the recommended dose goals.

Conclusion

The delivery of IMRT and volumetric-modulated arc therapy (VMAT) plans was verified to correspond well with calculated dose distributions for different pathologies. We found the Delta 4 device is accurate and reproducible. Although Delta4 appears to be a straightforward device for measuring dose and allows measure in real-time dosimetry QA, it is a complex device and careful quality control is required before its use.

Information

Type
Technical Note
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 Delta4 phantom set up at isocenter in measurement position.

Figure 1

Table 1 Average γ-passing rate, dose deviation and distance to agreement of different pathologies. Indicated statistical error uncertainty is standard deviation

Figure 2

Figure 2 A screenshot from the Delta4 software showing the result of seven beams intensity-modulated radiation therapy (IMRT) delivery for a nasopharynx case. The upper panel shows the absolute dose measured in two diode arrays in 3D with color coding denoting the dose. The lower panel shows three histograms: dose deviation from the planned dose, distance to agreement from the planned dose and gamma index distribution.

Figure 3

Figure 3 A screenshot from the Delta4 software showing seven beams intensity-modulated radiation therapy (IMRT) delivery. The upper panel shows the absolute dose difference in two diode arrays in 2D with color coding denoting the dose. The lower panel shows two graphs: dose profiles along X- and Y-axis.