Hostname: page-component-74656f9884-2dmjx Total loading time: 0 Render date: 2026-11-08T14:20:04.450Z Has data issue: false hasContentIssue false

Thermal Stress and Point-of-Care Testing Performance: Suitability of Glucose Test Strips and Blood Gas Cartridges for Disaster Response

Published online by Cambridge University Press:  08 April 2013

Rights & Permissions [Opens in a new window]

Abstract

Objective: Point-of-care testing (POCT) devices are deployed in the field for emergency on-site testing under a wide range of environmental conditions. Our objective was to evaluate the performance of glucose meter test strips and handheld blood gas analyzer cartridges following thermal stresses that simulate field conditions.

Methods: We evaluated electrochemical and spectrophotometric glucose meter systems and a handheld blood gas analyzer. Glucose test strips were cold-stressed (–21°C) and heat-stressed (40°C) for up to 4 weeks. Blood gas cartridges were stressed at –21°C, 2°C, and 40°C for up to 72 hours. Test strip and cartridge performance was evaluated using aqueous quality control solutions. Results were compared with those obtained with unstressed POCT strips and cartridges.

Results: Heated glucose test strips and blood gas cartridges yielded elevated results. Frozen test strips and cooled cartridges yielded depressed glucose and blood gas results, respectively. Frozen cartridges failed.

Conclusions: The performance of glucose test strips and blood gas cartridges was affected adversely by thermal stresses. Heating generated elevated results, and cooling depressed results. Disaster medical assistance teams and emergency medical responders should be aware of these risks. Field POCT devices must be robust to withstand adverse conditions. We recommend that industry produce POCT devices and reagents suitable for disaster medical assistance teams. (Disaster Med Public Health Preparedness. 2009;3:13–17)

Information

Type
Original Research and Critical Analysis
Copyright
Copyright © Society for Disaster Medicine and Public Health, Inc. 2009
Figure 0

FIGURE 1 Effects of thermal stress on glucose test strips. Mean differences in glucose results obtained with heat-stressed test strips relative to control for the 3 glucose meter systems (GMSs) are shown. Test strips were stressed at 40°C. Controls were stored and measured at room temperature (21°C). *P< 0.05. **P< 0.01. ***P< 0.001. Reprinted with permission from Knowledge Optimization, Davis, CA

Figure 1

FIGURE 2 Effects of freezing and thawing on glucose meter system (GMS) test strip results. This figure shows mean glucose difference between frozen or thawed test strips relative to control. Test strips were cold stressed at −21°C. Controls were stored and measured at room temperature (21°C). *P< 0.05. **P< 0.01. ***P< 0.001. Reprinted with permission from Knowledge Optimization, Davis, CA

Figure 2

FIGURE 3 Effects of heating and cooling of test cartridges on Po2 and Pco2 results obtained with a handheld blood gas analyzer. This figure shows the mean Po2 and Pco2 paired-differences between heated (40°C) or cooled (2°C) test cartridges and control (21°C). * P< 0.05. **P< 0.01. ***P< 0.001. Reprinted with permission from Knowledge Optimization, Davis, CA

Figure 3

TABLE 1 Storage and Operating Temperature (°C) Specifications