Evaluation of the surface dose for total body irradiation (TBI) technique with parallel-opposed anterior posterior geometry

Published online by Cambridge University Press:  09 March 2021

Hoseinnezhadzarghani Elham
Affiliation:
Department of Medical Physics and Medical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran
Geraily Ghazale*
Affiliation:
Department of Medical Physics and Medical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran Radiotherapy Oncology Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
Sarvin Sarmadi
Affiliation:
Department of Orthodontics, School of Dentistry, Tehran University of Medical Science, Tehran, Iran
*
Author for correspondence: Geraily Ghazale, Department of Medical Physics and Medical Engineering, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran. E-mail: gh-geraily@sina.tums.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

Aim:

Total body irradiation (TBI) is an external radiotherapy technique in which the whole body including the superficial regions is required to receive the therapeutic dose. The purpose of this study is to evaluate the received surface dose during TBI technique.

Methods and materials:

The anterior/posterior (AP/PA) TBI was implemented with 18-MV photon beam at 312-cm treatment distance for human-like phantom. The GAFCHROMIC-EBT3 films were used for superficial dose measurements.

Results and discussion:

The percentage of surface-absorbed dose relative to the prescription point for 8 points of measurements was between 102·78–121·48% and 104·51–127·43% at 5 and 10 mm depth, respectively. In the chest wall region due to the presence of lung blocks, the absorbed dose was below the acceptable level, so an electron boost was required to increase the chest wall absorbed dose.

Conclusions:

According to the results, the implemented technique was able to deliver sufficient dose to the shallow surface of phantom’s body.

Information

Type
Technical Note
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. During total body irradiation (TBI) technique implementation, the source-surface distance (SSD) was set at 312 cm, and the gantry of the Linac rotated 90 degrees. The phantom was positioned on left/right side, and the lungs blocks were mounted to the beam spoiler at 5 cm distance from the prescription point (umbilicus).

Figure 1

Figure 2. Points of surface dose measurement during total body irradiation (TBI) technique.

Figure 2

Figure 3. Percentage surface dose at the depth of 5 and 10 mm at 8 anatomical selected points.

Figure 3

Figure 4. The transversal absorbed dose profile at the level of umbilicus (the absorbed dose values were normalised to the absorbed dose at prescription point). Percentage absorbed dose at 5 and 10 mm is marked with dashed lines.

Figure 4

Figure 5. The electron and photon lung blocks. It can be seen that the shape of the electron blocks is exactly match to the shape of the photon lung blocks.