Real-time biofluorescent particle counting compared to conventional air sampling for monitoring airborne contamination in orthopedic implant surgery

Published online by Cambridge University Press:  07 April 2025

Frans Stålfelt*
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
Josefin Seth Caous
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, SwedenDivision Materials and Production, Methodology, Textiles, and Medical Technology, RISE Research Institutes of Sweden, Gothenburg, Sweden
Karin Svensson Malchau
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden
Camilla Björn
Affiliation:
Division Materials and Production, Methodology, Textiles, and Medical Technology, RISE Research Institutes of Sweden, Gothenburg, Sweden
Maziar Mohaddes
Affiliation:
Department of Orthopedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, SwedenDepartment of Clinical Sciences Lund – Orthopedics, Lund University, Lund, SwedenDepartment Orthopedics, Hässleholms Hospital, Hässleholm, Sweden
*
Corresponding author: Frans Stålfelt; Email: frans.stalfelt@vgregion.se

Abstract

Background:

Surgical site infection (SSI) following orthopedic surgery is a complication associated with morbidity and economic burden. Transmission of airborne bacteria that settle into surgical wounds constitutes a risk factor for SSIs. However, monitoring microbial contamination inside operating rooms with conventional methods is resource and time-consuming.

Aim:

This study aimed to assess correlation between a biofluorescent particle counter (BFPC) and conventional air sampling, to enable real-time monitoring of airborne contamination. Additionally, the study aimed to analyze correlation between particles near the surgical site and particles 1 meter away, to evaluate the feasibility of distance-based measurements.

Methods:

Correlation analysis was conducted to compare colony-forming units (CFU) collected using a Sartorius MD8 air sampler with biofluorescent viable particles detected by BioTrak 9510-BD, both positioned near the surgical site. Additionally, correlation between particle counts measured by AeroTrak 6510, positioned 1 meter away, and total particle counts measured by the BioTrak near the surgical site was evaluated. Sampling took place in two operating rooms: one with turbulent mixed airflow (TMA) and one with unidirectional airflow (UDAF).

Results:

Negligible to low correlation between biofluorescent particles and CFU was observed, both in UDAF (n = 100) and TMA (n = 22). However, strong correlation was found between BFPC and particle counter measurements of total numbers of particles (Rp = 0.634–0.769, P < .001).

Conclusion:

While BFPCs offer real-time monitoring of airborne contamination, their predictive ability for CFU levels remains uncertain. Yet, the strong correlation between particles in the surgical site and particles measured 1 meter away suggests feasibility to conduct future studies with larger cohorts.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2025. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Figure 1.A) Instrumental setup showing the biofluorescent particle counter probe alongside the tube and mouthpiece of the Sartorius MD8 air sampler, equipped with a gelatin membrane filter. B) Schematic of the instrument placement inside the operation room (OR) with vertical unidirectional airflow (UDAF) ventilation. The circled area indicates the aerial safe zone generated by the vertical UDAF. Similar instrumental placement was used in the turbulent mixed airflow ventilated OR, with the exception that no additional particle counter was used.

Figure 1

Table 1.Descriptive statistics of colony-forming unit (CFU) and logarithmic particle levels of all measured particle sizes

Figure 2

Table 2.Correlation coefficients in unidirectional airflow (UDAF)-ventilated operation room (OR) (n = 100) and turbulent mixed airflow (TMA)-ventilated OR (n = 22) between colony-forming unit (CFU) and BFP and particle counter (PC), and between total particle count measured by the biofluorescent particle counter (BFPC) and the PC

Figure 3

Figure 2.Scatter plot of total count of particles (1 µm) measured with a biofluorescent particle counter positioned close to the surgical site and with a conventional particle counter, positioned approximately 1m from the surgical site in a unidirectional airflow-ventilated OR.

Supplementary material: File

Stålfelt et al. supplementary material

Stålfelt et al. supplementary material
Download Stålfelt et al. supplementary material(File)
File 20 KB