Magnetic resonance imaging respirators: a randomized crossover trial to assess respiratory protection, usability, and comfort

Published online by Cambridge University Press:  29 April 2024

Daryl Lindsay Williams
Affiliation:
Department of Anaesthesia and Pain Management, Royal Melbourne Hospital, Parkville, VIC, AustraliaUniversity of Melbourne, Parkville, VIC, Australia
Benjamin Kave
Affiliation:
Department of Anaesthesia and Pain Management, Royal Melbourne Hospital, Parkville, VIC, AustraliaUniversity of Melbourne, Parkville, VIC, Australia
Charles Bodas
Affiliation:
Respiratory Protection Program, Royal Melbourne Hospital, Parkville, VIC, Australia
Megan Roberts
Affiliation:
Respiratory Protection Program, Royal Melbourne Hospital, Parkville, VIC, Australia
Irene Ng*
Affiliation:
Department of Anaesthesia and Pain Management, Royal Melbourne Hospital, Parkville, VIC, AustraliaUniversity of Melbourne, Parkville, VIC, Australia
*
Corresponding author: Irene Ng; Email: Irene.Ng@mh.org.au

Abstract

Objective:

Many available facepiece filtering respirators contain ferromagnetic components, which may cause significant problems in the magnetic resonance imaging (MRI) environment. We conducted a randomized crossover trial to assess the effectiveness, usability, and comfort of 3 types of respirators, judged to be “conditionally MRI safe” with an aluminum nosepiece (Halyard 46727 duckbill-type respirators and Care Essentials MSK-002 bifold cup-type respirators) or “MRI safe” completely metal free (Eagle AG2200 semirigid cup-type respirators).

Design and setting:

We recruited 120 participants to undergo a quantitative fit test (QNFT) on each of the 3 respirators in a randomized order. Participants then completed a usability and comfort assessment of each respirator.

Results:

There were significant differences in the QNFT pass rates (51% for Halyard 46727, 73% for Care Essentials MSK-002, and 86% for Eagle AG2200, P < .001). The first-time fit test pass rate and overall fit factor were significantly higher for Eagle AG2200 compared with the other 2 respirators. Eagle AG2200 scored the lowest ratings in the ease of use and overall comfort. There were no significant differences in other modalities, including the seal rating, breathability, firmness, and overall assessment.

Conclusions:

Our study supports the utility of the Eagle AG2200 and Care Essentials MSK-002 respirators for healthcare professionals working in an MRI environment, based on their high QNFT pass rates and reasonably good overall usability and comfort scores. Eagle AG2200 is unique because of its metal-free construction. However, its comparatively lower usability and comfort ratings raise questions about practicality, which may be improved by greater user training.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - SA
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike licence (http://creativecommons.org/licenses/by-nc-sa/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the same Creative Commons licence is used to distribute the re-used or adapted article and the original article is properly cited. The written permission of Cambridge University Press must be obtained prior to any commercial use.
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Figure 1.CONSORT diagram.

Figure 1

Table 1.Participants’ baseline characteristics. Values are expressed as mean+/– SD, number (percentage), and median (IQR [range])

Figure 2

Table 2.Quantitative fit test results of the 3 types of respirators: Halyard 46727 (Halyard), Care Essentials MSK-002 (MSK), and Eagle AG2200 (Eagle) N95/P2 respirators. Values are expressed as number (percentage), median (IQR [range])

Figure 3

Figure 2.Violin plots to show the ease of use, seal rating, and breathability among the 3 types of respirators: Halyard 46727 (Halyard), Care Essentials MSK-002 (MSK) and Eagle AG2200 (Eagle) N95/P2 respirators. *Friedman’s test shows significant differences among the 3 groups (P < .001). Pairwise comparisons show significant differences between Eagle and Halyard (P < .001) and between Eagle and MSK (P < .001) in the ease-of-use rating. (Violin plot is a combination of a box plot, which shows the summary statistics, and a kernel density plot, which shows the distribution of the data).

Figure 4

Figure 3.Violin plot to show the overall comfort rating for the 3 types of respirators: Halyard 46727 (Halyard), Care Essentials MSK-002 (MSK), and Eagle AG2200 (Eagle) N95/P2 respirators. *Friedman’s test shows a significant difference among the 3 groups (P = .007). Pairwise comparisons show a significant difference between Eagle and MSK (P < .012) only.

Figure 5

Table 3.Participants’ firmness of fit ratings for each of the three types of respirators: Halyard 46727 (Halyard), Care Essentials MSK-002 (MSK), and Eagle AG2200 (Eagle) N95/P2 respirators