Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-10T05:43:48.981Z Has data issue: false hasContentIssue false

Partially open stent after balloon catheter iatrogenic perforation. How to solve this complication using a novel technique? The Mij-Her technique

Published online by Cambridge University Press:  05 May 2022

Roberto Mijangos-Vázquez*
Affiliation:
Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Tuxtla Gutiérrez, Chiapas, México
Rogelio Hernández-Reyes
Affiliation:
Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Tuxtla Gutiérrez, Chiapas, México
*
Author for correspondence: R. Mijangos-Vázquez, Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Blvd. SS Juan Pablo II y Blvd, Antonio Pariente Algarín S/N, PC. 29070, Tuxtla Gutiérrez, Chiapas, México. Tel: (+52-961) 6170700. E-mail: dr.rmijangos@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Complications related to pulmonary artery stenting are potentially life-threatening. We reported a novel technique of how to achieve the introduction of a partial dilated stent into a long sheath using a snare in the event of a iatrogenic perforation of a balloon catheter.

There are no publications of similar techniques describing successful resolution of this type of complication.

Information

Type
Brief Report
Copyright
© The Author(s), 2022. Published by Cambridge University Press
Figure 0

Figure 1. (a) Snare crossing across the lumens of the balloon catheter. The snare is inside a multipurpose catheter. (b) Snare before crossing the haemostasis valve (arrow) of the Mullins-type sheath. (c) Snare inside the Mullins-type sheath (in the same lumen) of the balloon catheter (arrowhead). (d) Fluoroscopy. Snare exiting Mullins-type sheath and across the stiff wire and the balloon catheter (arrow). (e) Fluoroscopy. Snare entrapment and compression of the proximal edge of the partial dilated stent (arrowhead). (f) Fluoroscopy. Retrieving the stent back into the Mullins-type sheath (asterisk).

Figure 1

Figure 2. (a) Stent after removal using a snare. The stent is deformed in its proximal edge (arrow). (b) Stent after manual repair (arrowhead).