Single Multi-snare® retrieval of a > 16 mm Occlutech Figulla II atrial septal defect occluder—a case series

Published online by Cambridge University Press:  20 July 2026

Rik De Wolf*
Affiliation:
Pediatric cardiology, UZ Brussel, Belgium Pediatric cardiology, UZ Gent, Belgium
Kristof Vandekerckhove
Affiliation:
Pediatric cardiology, UZ Gent, Belgium
Joseph Panzer
Affiliation:
Pediatric cardiology, UZ Gent, Belgium
*
Corresponding author: Rik De Wolf; Email: rik.dewolf@uzbrussel.be
Rights & Permissions [Opens in a new window]

Abstract

Transcatheter closure is the preferred treatment for type 2 atrial septal defects. Device retrieval for embolisation or malpositioning is rarely necessary and can often be performed percutaneously. However, transcatheter retrieval of Occlutech Figulla Flex II atrial septal defects occluders >16 mm can be very difficult, and it has even been stated to be impossible in the literature. Its right atrial hub is ball shaped so that snare devices could slip off when retrieving a large device into a sheath, necessitating the use of multiple snare or other retrieval devices. We present three paediatric cases (four procedures in total) of successful > 16 mm Occlutech atrial septal defect occluder retrieval using a single Multi-Snare® device.

Information

Type
Case Report
Copyright
© The Author(s), 2026. Published by Cambridge University Press
Figure 0

Figure 1. (a) Atrial septal defect 2 and 3D sizing showing an oval-shaped 28 x 17 mm defect. (b) Anteroposterior angiographic view of the embolised atrial septal defects device to the right ventricle and the Lunderquist wire in the right pulmonary artery for support of the Multi-snare. (c) Lateral angiographic view of the anteriorly pointing right atrial hub of the device, caught by the Multi-snare.

Figure 1

Figure 2. (a) Echocardiographic image of the atrial septal defect size. (b) Echocardiographic image of the displaced device towards the mitral valve after release.

Figure 2

Figure 3. (a) Echocardiographic image of the malpositioned device with a significant space between the two discs (arrow). (b) Radiographic image of the malpositioned device being snared with the Multi-snare.

Supplementary material: File

De Wolf et al. supplementary material 1

De Wolf et al. supplementary material
Download De Wolf et al. supplementary material 1(File)
File 23.2 MB
Supplementary material: File

De Wolf et al. supplementary material 2

De Wolf et al. supplementary material
Download De Wolf et al. supplementary material 2(File)
File 10.1 MB