Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-08T10:57:57.823Z Has data issue: false hasContentIssue false

Feasibility of customised unipolar conversion using bipolar temporary pacing wires in patients after surgical repair of congenital heart disease

Published online by Cambridge University Press:  20 August 2013

Bjoern Peters*
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
Oliver Miera
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
Peter Ewert
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
Sevim Yilmaz
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
Felix Berger
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
Boris Schmitt
Affiliation:
Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Berlin, Germany
*
Correspondence to: B. Peters, MD, Department of Congenital Heart Disease and Pediatric Cardiology, Deutsches Herzzentrum Berlin, Augustenburger Platz 1, 13353 Berlin, Germany. Tel: +49 30 45932800; Fax: +49 30 45932900; E-mail: peters@dhzb.de
Rights & Permissions [Opens in a new window]

Abstract

Objective: Temporary pacing wires play a crucial role in the diagnosis and therapy of post-operative arrhythmia after surgery for congenital heart disease. At present, bipolar pacing wires are used in most institutions. In case of functional failure of these wires, a unipolar mode of stimulation and sensing should be theoretically possible as a rescue procedure. Methods: We tested the feasibility of the customised unipolar mode in 18 post-operative patients with congenital heart disease (age 9.2 ± 13.9 months, weight 6.3 ± 3.8 kg, and cardiopulmonary bypass time 70 ± 29 minutes). As there are two possible unipolar configurations, there are twice the number of testing parameters; of those, we compared sensing (mV) and pacing thresholds (V at 0.5 ms). Results: Atrial sensing was significantly better in the unipolar modes (p < 0.001, p < 0.003). The ventricular unipolar sensing did not differ significantly in the “better” of the two possible configurations from the bipolar values (p = 0.363). For the unipolar pacing thresholds, only the “better” unipolar configuration did not differ significantly from the bipolar measurements (atrial: p = 0.058, ventricular: p = 0.138). There was no exit block or undersensing. Conclusion: The results demonstrate that unipolar stimulation and sensing using bipolar epicardial temporary pacing wires is feasible. In the case of failure of bipolar temporary pacing wires, this modality represents an easy rescue measure that in such cases should always be considered.

Information

Type
Original Articles
Copyright
Copyright © Cambridge University Press 2013 
Figure 0

Figure 1 An external pacer (Pace 300, Osypka GmbH) with an atrial unipolar and ventricular bipolar configuration (example). Thus, for the atrium, only one pole of the bipolar electrode (white plug) is connected to the atrial channels as a cathode. A self-fabricated cable that is connected to a cutaneous patch electrode (see also Fig 2) serves as an anodal conductor. The ventricular lead is connected in a bipolar manner (both plugs in the ventricular channel).

Figure 1

Figure 2 For the unipolar configuration of the temporary pacing wires, a cutaneous defibrillation patch (DP-702 M Nikman™, Unomedical), placed on the left lateral thorax, serves as an anodal conductor. The patch is connected with two attached ECG electrodes (Ambu® Blue Sensor SP) and a self-fabricated cable to the corresponding channel of the external pacemaker.

Figure 2

Table 1 Testing values of the bipolar and unipolar modes in comparison, expressed as mean values with standard deviation and range.

Figure 3

Figure 3 (a–d) Comparison of bipolar and unipolar testing values using the Wilcoxon signed-rank test (boxplots with median and 25th/75th percentiles).