Evaluation of N95 respirator ultraviolet decontamination and clinical reuse with quantitative fit testing

Published online by Cambridge University Press:  03 May 2021

Gregory K. Wanner*
Affiliation:
Department of Emergency Medicine, ChristianaCare, Newark, Delaware
Douglas Ader
Affiliation:
Department of Emergency Medicine, ChristianaCare, Newark, Delaware
Richard Caplan
Affiliation:
Value Institute, ChristianaCare, Newark, Delaware
Amit S. Padaki
Affiliation:
Department of Emergency Medicine, ChristianaCare, Newark, Delaware
Debra Ravert
Affiliation:
Department of Emergency Medicine, ChristianaCare, Newark, Delaware
Marci Drees
Affiliation:
Departments of Medicine, Quality and Patient Safety, ChristianaCare, Newark, Delaware
*
Author for correspondence: Gregory K. Wanner, E-mail: gregory.wanner@christianacare.org
Rights & Permissions [Opens in a new window]

Abstract

The supply of N95 respirators has been severely strained by the coronavirus disease 2019 (COVID-19) pandemic. We used quantitative fit-testing to evaluate 16 participants and 45 respirators through up to 4 rounds of ultraviolet decontamination and clinical reuse. The mean fit-test failure rate was 29.7%, and the probability of failure increased through N95 reuse.

Information

Type
Concise Communication
Copyright
© The Author(s), 2021. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Table 1. Demographics and N95 Fit-Testing Results

Figure 1

Fig. 1. Kaplan-Meier curve showing probability of N95 fit-test passing based on 64 fit tests and 17 imputed data points. The gray shaded area represents 95% confidence intervals.