Polycrystalline materials analysis using the Maia pixelated energy-dispersive X-ray area detector

Published online by Cambridge University Press:  26 September 2017

Henry J Kirkwood
Affiliation:
ARC Centre of Excellence in Advanced Molecular Imaging, Department of Chemistry and Physics, La Trobe Institute for Molecular Science, La Trobe University, VIC 3086, Australia
Martin D de Jonge
Affiliation:
Australian Synchrotron, 800 Blackburn Road, Clayton, VIC 3168, Australia
Daryl L Howard
Affiliation:
Australian Synchrotron, 800 Blackburn Road, Clayton, VIC 3168, Australia
Chris G Ryan
Affiliation:
CSIRO Earth Sciences and Resource Engineering, Kensington, WA 6151, Australia
Grant van Riessen
Affiliation:
Department of Chemistry and Physics, La Trobe Institute for Molecular Science, La Trobe University, VIC 3086, Australia
Felix Hofmann
Affiliation:
Department of Engineering Science, University of Oxford, Parks Road, Oxford OX1 3PJ, UK
Matthew R Rowles
Affiliation:
Department of Physics and Astronomy, Curtin University, Perth, WA 6845, Australia
Anna M Paradowska
Affiliation:
Australian Centre for Neutron Scattering, Australian Nuclear Science and Technology Organisation, NSW 2234, Australia
Brian Abbey*
Affiliation:
ARC Centre of Excellence in Advanced Molecular Imaging, Department of Chemistry and Physics, La Trobe Institute for Molecular Science, La Trobe University, VIC 3086, Australia
*
a)Author to whom correspondence should be addressed. Electronic mail: b.abbey@latrobe.edu.au

Abstract

Elemental, chemical, and structural analysis of polycrystalline materials at the micron scale is frequently carried out using microfocused synchrotron X-ray beams, sometimes on multiple instruments. The Maia pixelated energy-dispersive X-ray area detector enables the simultaneous collection of X-ray fluorescence (XRF) and diffraction because of the relatively large solid angle and number of pixels when compared with other systems. The large solid angle also permits extraction of surface topography because of changes in self-absorption. This work demonstrates the capability of the Maia detector for simultaneous measurement of XRF and diffraction for mapping the short- and long-range order across the grain structure in a Ni polycrystalline foil.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2017 
Figure 0

Figure 1. (Colour online) The experimental setup at the XFM beamline. The incident beam energy was selected by the double-crystal monochromators (DCM) before being focused by the KB mirrors and passing through the center of the Maia detector and onto the polycrystalline nickel foil specimen.

Figure 1

Figure 2. Geometry for describing the spatial variations in XRF intensity due to self-absorption as described in the text. The beam is incident in the positive z-direction. Changes in the sample surface orientation lead to a change in self-absorption, which is seen in the detector plane as a change in the spatial XRF intensity.

Figure 2

Figure 3. (Colour online) (a) The normalized spectra measured in two Maia detector elements. Bragg diffraction was detected with a peak at approximately 9.3 keV in one of the pixels. The peaks at approximately 7.8 and 8.6 keV are Ni fluorescence peaks as labeled with the emission type. The Ni XRF signal is much higher than the diffraction signal. (b) The XANES spectra with and without a Bragg reflection, the height of the peak was easily detected above the XRF signal.

Figure 3

Figure 4. Example of processed (a) XAFS and (b) EXAFS spectra from the Ni foil experiment.

Figure 4

Figure 5. (a) The relative height of the Ni surface over the scan region. The scale bar is 20 µm and the black box shows the location of the hexagonal grain used for the subsequent topographic and diffraction analysis. (b) Surface relief map of the grain used in the diffraction analysis. The color scale shows the surface height relative to the center of the hexagonally shaped grain determined by XRF self-absorption. (c) Scanning electron microscope image of the same region as shown in (b).

Figure 5

Figure 6. (Colour online) (a) The magnitude of the gradient of the peak energy. The scale bar in the bottom left corner is 10 µm. (b) The FWHM of the peaks for the four measured reflections from the hexagonal grain, the colorbar scale is in eV. Reflections 3 and 4 are from twinned crystals and appear to run parallel to the directions of the expected slip planes for FCC crystals.

Supplementary material: PDF

Kirkwood et al supplementary material 1

Supplementary Figure

Download Kirkwood et al supplementary material 1(PDF)
PDF 357.5 KB