A customised laryngeal stent for glottic airway reconstruction

Published online by Cambridge University Press:  10 December 2021

A A Joshi
Affiliation:
Department of ENT, Lokmanya Tilak Municipal Medical College and General Hospital, Mumbai, India
S R Agrawal*
Affiliation:
Department of ENT, Lokmanya Tilak Municipal Medical College and General Hospital, Mumbai, India
R A Bradoo
Affiliation:
Department of ENT, Lokmanya Tilak Municipal Medical College and General Hospital, Mumbai, India
*
Author for correspondence: Dr S Agrawal, Department of ENT, Room no. 441, 4th floor, College building, Lokmanya Tilak Municipal Medical College & General Hospital, Sion, Mumbai 400022, India E-mail: drshailesh3187@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objective

Laryngotracheal reconstructive surgery very often requires placement of laryngeal stents, but most of the available stents either do not conform to the shape of the glottis or are open at the upper end, causing significant tissue reaction and aspiration. LT-Mold is an ideal stent in this respect, but it is not readily available.

Case report

The authors of this study fashioned a laryngeal stent out of a readily available Montgomery T-tube in such a way that it had a closed upper end and was well-conformed to the architecture of the larynx. This stent was used in a case of posterior expansion laryngoplasty, and post-operatively the patient had a patent airway with good voice and no aspiration.

Conclusion

A customised laryngeal stent is well tolerated and helps to solve complicated cases of posterior glottic stenosis.

Information

Type
Clinical Records
Copyright
Copyright © The Author(s), 2021. Published by Cambridge University Press on behalf of J.L.O. (1984) LIMITED.
Figure 0

Fig. 1. Vertical limb of Montgomery T-tube that was cut to design a laryngeal stent. The distance between the vocal fold and superior margin of tracheostoma is measured on vertical limb of the T-tube (i.e. 3 cm). A = distance between the vocal fold (anterior commissure) and the superior margin of the tracheostoma measured and marked as A from the distal end of tube.

Figure 1

Fig. 2. Stepwise designing of a new laryngeal stent using a silicon Montgomery T-tube. (a) Distance between the vocal fold and the superior margin of the tracheostoma measured from the distal end of the tube, marked as A. (b) Horizontal line drawn at 45 degrees from point A and labelled as B1 and B2. Anterior half of tube sectioned vertically at its proximal end. (c) Posterior flap trimmed in triangle shape to fold anteriorly (point C). (d) Triangular posterior flap sutured anteriorly.

Figure 2

Fig. 3. (a) Laryngofissure with placement of posterior costal cartilage graft. (b) New laryngeal stent placed over posterior graft and prolene stitch secured at the level of anterior commissure.

Figure 3

Fig. 4. Endoscopic images showing: (a) intra-operative confirmation of stent, (b) endoscopic view of stent at one-month post-surgery, (c) uncapping of stent, (d) cutting of stay sutures, (e) removal of stent and (f) after stent removal.

Figure 4

Fig. 5. Endoscopic images showing: (a) pre-operative endoscopic view of glottis stenosis, (b) immediately after stent removal and (c) two weeks after stent removal showing expanded glottis area.