A dosimetric evaluation of a novel technique using abutted radiation fields for myeloablative total body irradiation

Published online by Cambridge University Press:  14 December 2020

Arun Chairmadurai*
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India Amity Centre for Radiation Biology, Amity University, Noida, India
Raghul Ramiya Jayabalan
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India
Thirumal Mani
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India
Abhishek Gulia
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India
Hari Mohan Agrawal
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India
Rekha Arya
Affiliation:
Department of Radiation Oncology, Jaypee Hospital, Noida, India
Esha Kaul
Affiliation:
Department of Hemato-Oncology (BMT), Jaypee Hospital, Noida, India
Nivedita Dhingra
Affiliation:
Department of Hemato-Oncology (BMT), Jaypee Hospital, Noida, India
*
Author for correspondence: Arun Chairmadurai, Department of Radiation Oncology, Jaypee Hospital, Noida, 201304, India. Tel: +91 9958198500. E-mail: arun2007ceg@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Background:

The present study reports myeloablative total body irradiation (TBI) on an isocentrically mounted linac by laying the patient on the floor and management of abutting radiation fields and partial shielding of lungs. Dosimetrical efficacy of this novel technique was evaluated.

Materials and methods:

In this retrospective study, dosimetrical parameters from TBI plans on whole-body CT scans of 46 patients were analysed. The prescribed dose to TBI was 12 Gy in six fractions delivered over a period of 3 days for myeloablative conditioning. TrueBeam STx platform Linac (Varian Medical Systems Inc., Palo Alto, CA, USA) was used to deliver opposing fields. Radiation fields were abutted to form a single large field using an arithmetic formula at source-to-skin-distance of 210 cm.

Results:

Discrepancies in dose calculated by treatment planning system were within 1·6% accuracy, and dose profile at the junction of abutting radiation fields was reproduced within 3·0% accuracy. The real treatment time for each patient was ~30 minutes/fraction. Monitor unit was weighted for multiple sub-fields to achieve dose homogeneity within 5·0% throughout the whole body, and the mean dose to lung was ≤10 Gy.

Conclusion:

Our abutting radiation field technique for myeloablative TBI is feasible in any existing linac bunker. ‘Island-blocking’ is feasible in this technique using multi-leaf collimator. This technique is cost-effective as it does not require any costly equipment than the readily available equipment in any radiotherapy facility. In general, TBI requires laborious planning procedures and spacious linac bunkers; this novel technique has the potential to change previously held notions.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Description of abutting radiation fields for myeloablative TBI by laying the patient on the floor.

Figure 1

Figure 2. Formulation of abutting radiation fields for myeloablative TBI.

Figure 2

Figure 3. Dose distribution calculated by TPS for abutting radiation fields and the dose profile at 5 cm depth on a coffin-box phantom.

Figure 3

Figure 4. Dosimetrical set-up to acquire auto-radiograph at the junction of abutting radiation fields: (a) gantry tilt—0° and (b) gantry tilt—15°.

Figure 4

Figure 5. Treatment fields for myeloablative TBI.

Figure 5

Table 1. Treatment plan parameters of abutting radiation fields used in paediatric and adult patients for myeloablative TBI

Figure 6

Figure 6. Sub-fields to boost the dose which was under-dosed from thoracic field (Figure 5a) to deliver a mean dose of 10 Gy to lungs.

Figure 7

Figure 7. Visual verification of lung shield with the optical field from treatment head by an expert radiologist.

Figure 8

Table 2. Point dose validation between measured and TPS calculated doses/1,000 MUs at 209 cm—SSD

Figure 9

Figure 8. (a) Illuminated auto-radiographs of Gafchromic films on X-ray view box (M/s. Bio-x) at 6500 LUX. (b) Comparison of intensity profile from auto-radiograph and dose profile calculated from TPS.

Figure 10

Table 3. Summary of dose received by PTV, brain, kidneys and lungs

Figure 11

Table 4. Influential parameters to reduce treatment time in myeloablative TBI

Figure 12

Table 5. Dose received by 15 patients who underwent myeloablative TBI in our institute

Figure 13

Figure 9. (a) Three-dimensional dose distribution of myeloablative TBI. (b) Dose distribution in frontal view. (c) ‘Island-block’ for shielding lungs using HD120TMMLC.