Introduction
Neural prosthetic devices, which (re)connect the brain with the outside world, promise to be useful for many clinical applications.Reference Hochberg, Serruya, Friehs, Mukand, Saleh, Caplan, Branner, Chen, Penn and Donoghue1–Reference Schwartz, Cui, Weber and Moran9 Microelectrodes that are inserted into the brain and left to reside in the neuronal cell body layers of the cortex are the key element to this technology.Reference Ward, Rajdev, Ellison and Irazoqui3 Such electrodes can record the activity of individual or small populations of neurons for a short time after implantation,Reference Schwartz8, Reference Buzsáki10, Reference Nicolelis11 but it has proven difficult to record neural signals over long periods.Reference Ward, Rajdev, Ellison and Irazoqui3 This remains one of the largest hurdles in translating research successes in this field into clinical implementation and standard of care.Reference Ward, Rajdev, Ellison and Irazoqui3, Reference Leach, Achyuta and Murthy6, Reference McConnell, Rees, Levey, Gutekunst, Gross and Bellamkonda12–Reference Williams, Rennaker and Kipke14
While the dominant mechanism is still being debated, it is well established that the recording ability of intracortical microelectrodes is related to the proximity of viable neurons and the characteristics of non-neural tissue between the electrode and neurons.Reference Buzsáki10, Reference Biran, Martin and Tresco15 The most widely accepted hypothesis for electrode failure is the development of an encapsulating scar at the electrode/tissue interface.Reference Ward, Rajdev, Ellison and Irazoqui3, Reference McConnell, Rees, Levey, Gutekunst, Gross and Bellamkonda12, Reference Polikov, Block, Fellous, Hong and Reichert13, Reference Biran, Martin and Tresco15–Reference Szarowski, Andersen, Retterer, Spence, Isaacson, Craighead, Turner and Shain17 Insertion-related damage is practically unavoidable and causes an acute tissue response; this effect can be minimized by reducing the insertion trauma.Reference Johnson, Kao and Kipke18–Reference Sharp, Panchawagh, Ortega, Artale, Richardson-Burns, Finch, Gall, Mahajan and Restrepo20 For example, Sharp and colleagues have shown that implanting the device more slowly can reduce initial scar formation.Reference Sharp, Panchawagh, Ortega, Artale, Richardson-Burns, Finch, Gall, Mahajan and Restrepo20 Interestingly, however, several studies showed that the chronic response is independent of the acute reaction induced by surgical trauma.Reference Biran, Martin and Tresco15, Reference Bjornsson, Oh, Al-Kofahi, Lim, Smith, Turner, De, Roysam, Shain and Kim19 Such studies have revealed two important phenomena: (1) the formation of a dense encapsulating scar, which eventually encapsulates the electrodes, regardless of the materials used; and (2) the formation of a neuron-free dead zone surrounding the implants.Reference Ward, Rajdev, Ellison and Irazoqui3, Reference Biran, Martin and Tresco15, Reference He, Bellamkonda and Reichert21–Reference Tresco and Winslow23 Numerous factors are thought to contribute to scar formation, including the mechanical properties mismatch between stiff electrodes and the soft cortical tissue,Reference Kim, Hitchcock, Bridge and Tresco24–Reference McConnell, Schneider, Owens and Bellamkonda26 and a localized neurotoxic environment created by the presence of a non-removable foreign object.Reference Zhong and Bellamkonda22–Reference Kim, Hitchcock, Bridge and Tresco24 The aim of this article is to summarize the different approaches to stabilize the neural electrode/brain interface by focusing on the mechanical mismatch of the implant and tissue, and, as an example, highlight recent work on mechanically adaptive nanocomposites for neural interfacing.
Strategies for stabilizing the neural electrode/brain interface
Most experimental microelectrodes are made from metals, silicon, and/or ceramics (i.e., materials that are much stiffer [Young’s modulus, E∼200 GPa] than cortical brain tissue [E∼0.1–6 kPa]). These rigid materials generally enable facile electrode insertion and have provided many of the major advances in the field.Reference Ward, Rajdev, Ellison and Irazoqui3, Reference Donoghue5, Reference Lebedev and Nicolelis27 However, once implanted, they exert forces on the surrounding tissue, since pulsing, respiration-related, and everyday motions cause movements of the brain relative to the device.Reference Gilletti and Muthuswamy28 The resulting shear forces scale with the stiffness difference between tissue and implant.Reference Subbaroyan, Martin and Kipke29 For example, electrodes placed into cortical tissue without anchoring to the skull (referred to as “un-tethered”) exhibit smaller differential motions than identical electrodes that are anchored to the skull (referred to as “tethered”) and cause less inflammatory-mediated scarring.Reference Biran, Martin and Tresco30 Mechanical modeling predicts that stresses induced in the surrounding tissue are reduced if the electrode’s stiffness is lowered,Reference Subbaroyan, Martin and Kipke29 and better tissue integration will minimize strain.Reference Lee, Bellamkonda, Sun and Levenston25 Therefore, the effects of the mechanical mismatch are thought to play a significant role in the cell-mediated inflammatory response affecting the microelectrode/cortical tissue interface.Reference Leach, Achyuta and Murthy6, Reference Polikov, Tresco and Reichert16, Reference He, Bellamkonda and Reichert21, Reference Williams, Hippensteel, Dilgen, Shain and Kipke31
Several strategies are being investigated to improve microelectrode biocompatibility and to minimize inflammatory responses,Reference Leach, Achyuta and Murthy6, Reference He, Bellamkonda and Reichert21 typically by promoting a more intimate interface between the neurons and the electrode surface. Several therapeutic approaches have demonstrated temporal attenuation of certain aspects of foreign body encapsulation and/or neuronal dieback at the device interface,Reference Szarowski, Andersen, Retterer, Spence, Isaacson, Craighead, Turner and Shain17, Reference Kim and Martin32–Reference He, McConnell, Schneider and Bellamkonda39 but a clinically viable approach for long-term recording without the risk of significant side effects has yet to be developed. Several groups investigated the effects of mechanical mismatch between microelectrodes and cortical tissue by fabricating microelectrodes and, in some cases, auxiliary coatings from a variety of “off-the-shelf” polymeric materials.Reference Rousche, Pellinen, Pivin, Williams, Vetter and Kipke40–Reference Lu and Shoichet46 In general, these materials have yet to demonstrate success in attenuating scar formation or improving neural recordings, perhaps because most of the materials employed had Young’s moduli that were still 10Reference Leach, Achyuta and Murthy6 times higher than that of the brain.Reference Takeuchi, Ziegler, Yoshida, Mabuchi and Suzuki42, Reference Wester, Lee and LaPlaca43, Reference Lee, He, Clement, Massia and Kim47–Reference Takeuchi, Suzuki, Mabuchi and Fujita49 Implanting compliant electrodes is difficult,Reference Takeuchi, Ziegler, Yoshida, Mabuchi and Suzuki42, Reference Kozai and Kipke50–Reference Kee-Keun, He, Singh, Massia, Ehteshami, Kim and Raupp52 because soft devices buckle during (attempted) insertion into cortical tissue unless special measures are taken.Reference Kozai and Kipke50 Such special measures are often only applicable to single shank electrode systems and are difficult, if not impossible, to apply to arrayed electrode devices. Mechanical modeling reveals that a Young’s modulus of several GPa is required to allow for unassisted insertion of the electrode with a cross-sectional area of around a thousand μm2 into cortical brain tissue. On this basis, significant efforts were made to develop electrodes based on semi-flexible polymer substrates and “hybrids” between rigid silicon cores and soft hydrogel coatings (for a review, see Reference Reference Potter, Gui, Capadona and Bar-Cohen53).
Comparison of mechanically adaptive nanocomposites to other electrode materials
The concept of using mechanically adaptive nanocomposites as the basis for cortical electrodes potentially offers some advantages over the current generation of electrode materials. Perhaps most importantly, the soft modulus after implantation (ca. 10 MPa) is significantly lower than that of silicon or tungsten (∼100 GPa)Reference Lu, Franze, Seifert, Steinhauser, Kirchhoff, Wolburg, Guck, Janmey, Wei, Kas and Reichenbach54 or even that of other flexible polymers used in neural interfacing, such as polyimide,Reference Rousche, Pellinen, Pivin, Williams, Vetter and Kipke40 poly(benzocyclobutene),Reference Lee, He, Clement, Massia and Kim47 or poly(p-xylylenes) (parylene) (∼2–3 GPa).Reference Takeuchi, Ziegler, Yoshida, Mabuchi and Suzuki42, Reference Seymour and Kipke55 Furthermore, electrodes based on such flexible polymers require temporary stiffening to allow for implantation, something that is not required by an appropriate mechanically adaptable material (vide infra). Nonetheless, a number of innovative approaches have been explored to achieve temporary stiffening of a polymer-based electrode. For example, the coating of a polymer device with biodegradable rigid polymers such as poly(lactic-co-glycolic acid) (PLGA), deposition of a thin layer of silicon, or the use of a “carrier” to insert the electrode.Reference Kozai and Kipke50, Reference Kee-Keun, He, Singh, Massia, Ehteshami, Kim and Raupp52, Reference Lee, Singh, He and Massia56 More recently, Shain and co-workers reported a biodegradable terpolymer coating system with a modulus of around 400 MPa, which required a larger than normal cross-sectional area of the electrode to allow for reliable implantation.Reference Lewitus, Smith, Shain and Kohn57 While each of these approaches has been demonstrated to be useful for the insertion of simple single-shank electrodes, challenges still exist if they are to be applied to multi-shank probes with an integrated lead cable. Furthermore, the use of a “carrier” system (i.e., a rigid insertion device around the electrode, which is retracted after insertion) to deliver the microelectrode also increases the volume of damaged tissue. Several studies have suggested that the damage caused during insertion of the electrodes will eventually heal, and neurons will repopulate the damaged tissue (vide infra). However, to date, no such studies have been reported on “carrier” models, and it is still unclear if such carriers will compound the wound healing response with the addition of a foreign body reaction. Finally, the use of a degradable coating adds an additional level of complexity in evaluating the bio-response, as one must also consider how the degradation products may affect the inflammatory response in the surrounding tissue.
Another interesting approach, pioneered by Martin and co-workers, relies on the coating of rigid electrodes with hydrogels that can reach E valves lower than the current generation of mechanically adaptive nanocomposites.Reference Ludwig, Uram, Yang, Martin and Kipke58–Reference Xiao, Martin, Cui and Shenai63 The hydrogel is dehydrated prior to implantation to render it stiff and prevent it from being stripped off the electrode during insertion. Once inserted, the coating absorbs large amounts of water from the tissue, causing it to swell and form a very soft hydrogel. It is yet unclear if the hydrogels, which serve as a mechanical “buffer,” are capable of masking the mechanical strain on the tissue, or if the underlying bulk stiffness will play a significant role in host response. The extensive swelling of the hydrogel coating increases the volume occupied by the electrode after insertion and “pushes” the neurons away from the actual electrical contact. To address this point, recent approaches to bridge the gap between the electrode and the tissue involve the grafting of conductive polymers such as polypyrrole or poly(3,4-ethylenedioxythiophene) (PEDOT),Reference Kim, Abidian and Martin64, Reference Yang, Kim, Hendricks, Leach, Northey and Martin65 or bioactive peptides to promote neuronal ingrowth onto the surface of the electrodes.Reference Cui, Hetke, Wiler, Anderson and Martin66, Reference Cui, Lee, Raphael, Wiler, Hetke, Anderson and Martin67 Immunohistological results of in vivo testing of the hydrogel electrode system have yet to be reported, but recording studies suggest improved signal quality in short-term animal studies.Reference Ludwig, Uram, Yang, Martin and Kipke58, Reference Kim, Wiler, Anderson, Kipke and Martin59
Innovation through bioinspiration
We set out with the goal of creating a microelectrode that would be sufficiently stiff to be easily inserted into the brain, but would subsequently soften upon implantation to better match the stiffness of the cortex, thereby minimizing forces exerted on the tissue and attenuating inflammation.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68 To this end, we have developed a new class of bioinspired, chemoresponsive, mechanically adaptive polymer nanocomposites that can controllably and selectively be switched between stiff and compliant states.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68–Reference van den Berg, Capadona and Weder74 The design of these materials is based on the architecture and functionality of the sea cucumber dermis.Reference Trotter and Koob75–Reference Trotter, Tipper, Lyons-Levy, Chino, Heuer, Liu, Mrksich, Hodneland, Shannon-Dillmore, Koob, Koob-Emunds, Kadler and Holmes77 These invertebrates can rapidly and reversibly switch the stiffness of their skin (Figure 1a–b).Reference Motokawa78–Reference Szulgit and Shadwick81 This response is achieved through a nanocomposite design that relies on rigid collagen fibers, which reinforce a soft matrix. The stiffness of the tissue is regulated by controlling the interactions, and thereby stress transfer, between adjacent collagen fibrils. Our artificial nanocomposites mimic the sea cucumber’s general design and consist of a soft polymer matrix that is reinforced with rigid nanofibers (Figure 1c). The magnitude of the reinforcement depends mainly on the modulus of the nanofibers, their aspect ratio (ratio of length to width), and concentration in the matrix, as well as their interactions between each other and with the polymer matrix. In the “on state,” the nanofibers strongly interact with each other so that they form a load-bearing percolating network, leading to a high overall stiffness. Upon introduction of a chemical switch that turns off the interactions between the nanofibers, the load-bearing network is disassembled, and the material’s stiffness is reduced.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68, Reference Shanmuganathan, Capadona, Rowan and Weder72, Reference Hsu, Rowan and Weder82
Picture of the natural model, a sea cucumber, in (a) the relaxed soft state and (b) the stiffened state, and (c) a schematic of the architecture proposed for the sea cucumber dermis and of the retro-engineered biomimetic nanocomposites used for mechanically dynamic intracortical microelectrodes. Photos courtesy of F. Carpenter of Frc Photography.

Multiple generations of artificial, mechanically adaptive nanocomposites
To achieve the desired mechanical switching in nanofiber-containing nanocomposites, it is critical that the filler particles form a rigid reinforcing network within the polymer matrix, and their interactions can be tuned through a stimulus in a controlled manner.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68, Reference Shanmuganathan, Capadona, Rowan and Weder71, Reference Shanmuganathan, Capadona, Rowan and Weder72, Reference Eichhorn, Dufresne, Aranguren, Capadona, Rowan, Weder, Thielemans, Roman, Renneckar, Gindl, Weigel, Yano, Abe, Nogi, Mangalam, Simonsen, Benight, Bismarck and Berglund83 We chose to employ highly crystalline, bio-based cellulose nanofibers (also referred to as cellulose nanocrystals, nanowhiskers, or simply whiskers),Reference Eichhorn, Dufresne, Aranguren, Capadona, Rowan, Weder, Thielemans, Roman, Renneckar, Gindl, Weigel, Yano, Abe, Nogi, Mangalam, Simonsen, Benight, Bismarck and Berglund83 which, depending on the source, display a Young’s modulus of 100–150 GPa,Reference Sturcova, Davies and Eichorn84, Reference Rusli and Eichhorn85 an ultimate tensile strength of up to 10 GPa, and an aspect ratio of up to 100.Reference Hsieh, Yano, Nogi and Eichhorn86 Due to the abundance of surface hydroxyl groups, cellulose whiskers strongly interact with each other through hydrogen bonding, but exposure to water efficiently reduces whisker-whisker interactions on account of competitive hydrogen bonding; this feature was exploited to create water-responsive, mechanically adaptive materials. Nanocomposites, in which the cellulose whiskers form a percolating network in a polymer matrix, can be fabricated by combining the components in a hydrogen-bond-forming solvent so that the whisker-whisker interactions are “switched off.” Upon film casting and evaporation of the solvent, whisker interactions are switched on, and the whiskers assemble into a percolating network within the polymer matrix.
In our first generation of mechanically adaptive nanocomposites, cellulose whiskers isolated through acid hydrolysis of the mantles from tunicates—sessile sea creatures—were employed as the responsive filler.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68 These tunicate nanowhiskers (TNWs) feature a diameter of ∼20 nm, an aspect ratio of ∼70–100, and an on-axis tensile storage modulus E′ of 120–150 GPa. Their high aspect ratio and stiffness make TNWs a better filler than cellulose whiskers from other sources, but these other sources, including wood, hemp, flax, jute, ramie, bacteria, and cotton, are often technologically more viable (vide infra). TNWs were integrated into a rubbery ethylene oxide-epichlorohydrin (EO-EPI) copolymer matrix,Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68, Reference Capadona, van den Berg, Capadona, Schroeter, Rowan, Tyler and Weder70, Reference Schroers, Kokil and Weder87 which in its neat form displays a low tensile storage modulus (E′ = 0.3–3 MPa, depending on the comonomer ratio) and can be swelled slightly with water. The incorporation of TNWs into the EO-EPI matrix causes a dramatic stiffness increase.Reference Capadona, van den Berg, Capadona, Schroeter, Rowan, Tyler and Weder70E′ increased with increasing TNW content from 1.3 MPa for the neat polymer to 800 MPa at a TNW content of 17% v/v.
As designed, the nanocomposites exhibit a significant stiffness reduction upon exposure to water (Figure 2a). For example, E′ dropped from 800 to 20 MPa for the material comprising 17% v/v TNWs. This response was completely reversible; when the materials were dried, the original properties were restored.Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68, Reference Capadona, van den Berg, Capadona, Schroeter, Rowan, Tyler and Weder70 The mechanical properties of these EO-EPI/TNW nanocomposites (and all others discussed here) are well described by two mechanical models: the percolation model for the dry state and the Halpin-Kardos model for the wet state (Figure 2a). These models, which are discussed in detail elsewhere,Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68, Reference Takayanagi, Uemura and Minami88, Reference Ouali, Cavaillé and Pérez89 can be used to predict the mechanical on-off contrast of the adaptive nanocomposites as a function of composition. Quantitative insights into the stress-transfer mechanisms that determine the films stiffness were gained by Raman spectroscopy.Reference Rusli, Shanmuganathan, Rowan, Weder and Eichhorn90 A diagnostic Raman band, associated with the C-O ring stretching of the cellulose backbone, was used to quantify the local orientation of and the level of stress experienced by the cellulose whiskers. It was shown that the extent of stress-transfer is influenced by local orientation and connectivity of the cellulose whiskers; these features are governed by the processing conditions used to fabricate the materials.Reference Rusli, Shanmuganathan, Rowan, Weder and Eichhorn90, Reference Rusli, Shanmuganathan, Rowan, Weder and Eichhorn91
(a) Tensile storage moduli E′ of ethylene oxide-epichlorohydrin (EO-EPI)/tunicate whisker nanocomposites as a function of volume fraction of whiskers. The nanocomposites were conditioned by either drying in vacuum, equilibrium swelling in deionized water, or swelling to saturation in deionized water followed by re-drying in vacuum. The lines represent values predicted by the percolation (solid) and Halpin-Kardos (dotted) models. The arrow indicates changes in modulus and volume fraction of whiskers resulting from aqueous swelling of one selected sample (19% v/v tunicate whiskers). (b) Time-dependent modulus decrease of neat poly(vinyl acetate) (PVAc) and a 12.2% v/v PVAc/tunicate whisker nanocomposite upon immersion into artificial cerebrospinal fluid (ACSF) and increasing the temperature from 23°C to 37°C. Lines represent time required for temperature to increase from 23°C to 37°C and isothermal control at 37°C. Reprinted with permission from Reference Reference Capadona, Shanmuganathan, Tyler, Rowan and Weder68. ©2008, AAAS.
Unfortunately, the stiffness of EO-EPI/TNW nanocomposites (E′ of up to 800 MPa) is too low to allow for the fabrication of electrodes that can be inserted into the brain through the outermost protective membrane on the surface of the brain, the pia mater. The next generation of mechanically adaptive nanocomposites was therefore designed to exploit two complementary switching mechanisms. Poly(vinyl acetate) (PVAc), an amorphous polymer with a glass transition temperature (T g) around 42°C, was employed as the matrix and cellulose whiskers isolated from tunicates as the filler. Upon exposure to physiological conditions, the materials undergo a phase transition (water plasticizes the material and lowers T g to below physiological temperature), and the whisker network is disassembled. This design allowed us to create adaptive PVAc/TNW nanocomposites, which exhibit a mechanical contrast of three orders of magnitude between the dry state at room temperature (E′ = 5.1 GPa for a nanocomposite with 16.5% v/v TNWs) and the water-swollen state at 37°C (∼12 MPa).Reference Capadona, Shanmuganathan, Trittschuh, Seidel, Rowan and Weder69, Reference Shanmuganathan, Capadona, Rowan and Weder71Figure 2b, which shows the stiffness of a thin film of a PVAc/TNW nanocomposite with 12.2% v/v TNWs upon introduction into artificial cerebrospinal fluid (ACSF) at 37°C, reveals that most of the softening occurs within minutes (i.e., over a timeframe that appears to be very useful for the electrodes to be placed into neural tissue).
One perceived drawback of the PVAc/TNW nanocomposites is their rather significant water or ACSF take-up (∼70–90% w/w for materials with 16.5% v/v TNWs at 37°C). An excessive level of swelling can lead to delamination of the multilayer electrode structure and increases the trauma. This problem was alleviated by introducing cellulose nanowhiskers (CNWs) isolated from cotton into PVAc. The CNWs have a diameter of 10–20 nm, a length of 100–250 nm, an average aspect ratio of ∼10 (which is much lower than the aspect ratio of TNWs [70–150]), and a tensile storage modulus of ∼105 GPa.Reference Capadona, Shanmuganathan, Trittschuh, Seidel, Rowan and Weder69, Reference Rusli and Eichhorn92 The CNWs also feature a lower density of surface sulfate groups (∼31 mmol/kg) than TNWs (∼85 mmol/kg). These anionic sites, introduced during hydrolysis of the cellulose pulp with sulfuric acid, aid in their dispersion, but they are also thought to be responsible for the hygroscopic nature of the TNW-based nanocomposites. Gratifyingly, PVAc/CNW nanocomposites show dynamic mechanical properties that are close to those of PVAc/TNW nanocomposites, but their water take-up is greatly reduced (∼28% w/w for a material with 16.5% v/v CNWs at 37°C).Reference Shanmuganathan, Capadona, Rowan and Weder72
Mechanically adaptive CNW nanocomposites with a rubbery polyurethane matrix were also explored.Reference Mendez, Annamalai, Eichhorn, Rusli, Rowan, Foster and Weder93 These materials are characterized by high toughness, exhibit water-responsive, mechanically adaptive properties, and also display shape-memory behavior. In the water-swollen, softened state, tensile deformation of the nanocomposites causes uniaxial orientation of the CNWs, which is retained upon drying the materials. The resulting oriented network fixates a temporary shape, which due to the elastic nature of the polyurethane matrix relaxes to the original shape upon wetting. Thus, this latest generation of water-triggered mechanically adaptive materials not only permits the fabrication of cortical electrodes, which soften after implantation, but also allows one to pre-program a shape change to correspond to movements that could simplify electrode placement in the tissue, or even facilitate a secondary position of the electrode, after the foreign body response has stabilized.
Implant fabrication and in vivo deployment
With this family of mechanically adaptive, physiologically responsive materials in hand, model microprobes consisting of neat PVAc only (as reference) and 12.2% v/v PVAc/TNW nanocomposites were created (Figure 3), with the goal of implanting them into a living rodent to study the inflammatory response.Reference Harris, Hess, Rowan, Weder, Zorman, Tyler and Capadona94, Reference Hess, Capadona, Shanmuganathan, Hsu, Rowan, Weder, Tyler and Zorman95 This first required the development of microfabrication processes compatible with the properties (rheology, chemical, and thermal stability) of these nanocomposites to allow for the integration of electrode materials and additional insulation layers.Reference Hess, Capadona, Shanmuganathan, Hsu, Rowan, Weder, Tyler and Zorman95 Based on the process shown in Figure 4, micromachined probes (50–100 μm thick) were fabricated that incorporated parylene as an insulating/moisture barrier layer (1 μm) and Ti/Au contacts (50 nm Ti, 200 nm Au).
(a) Micromachined dogbone structure with lithographically defined Ti/Au pads and trace for evaluating mechanical properties; (b) laser-micromachined cortical probe with a lithographically defined Ti/Au electrode, interconnects, and parylene capping layers (cf. Figure 4, entry 10). In both cases, a 12.2% v/v poly(vinyl acetate) (PVAc)/tunicate whisker nanocomposite was used as the mechanically adaptive substrate (thickness 60 μm). Reprinted with permission from Reference Reference Hess, Capadona, Shanmuganathan, Hsu, Rowan, Weder, Tyler and Zorman95. ©2011, IOP Publishing.
Fabrication process used to integrate metal electrodes and parylene insulation layers onto a chemically sensitive poly(vinyl acetate) (PVAc)/tunicate whisker (TW) substrate. Reprinted with permission from Reference 95. ©2011, IOP Publishing.
The characterization of probes manufactured by this approach revealed that the parylene and gold layers, in spite of their intrinsic stiffness, only marginally contributed to the overall stiffness, due to their comparably small thickness. Gratifyingly, these electrodes exhibit impedances (95.0 ± 4.8 kΩ) that are comparable to those of Au electrodes (130 kΩ) on conventional substrates. Interestingly, the swelling of the nanocomposite-based electrodes was highly anisotropic, perhaps on account of processing-related anisotropy;Reference Eichhorn, Dufresne, Aranguren, Capadona, Rowan, Weder, Thielemans, Roman, Renneckar, Gindl, Weigel, Yano, Abe, Nogi, Mangalam, Simonsen, Benight, Bismarck and Berglund83 the expansion in the thickness dimension exceeded that of the in-plane direction by a factor of ∼10. This feature explains, at least in part, the absence of discernible deformation and the nonexistence of delamination between the nanocomposite and parylene in the probes when deployed in deionized water. In the case of dog bone structures, used exclusively for mechanical testing (thickness of ca. 50 μm), the Young’s modulus decreased from ca. 3400 to ca. 20 MPa within only 300 seconds and displayed essentially the same mechanical contrast as the nanocomposites alone.
Figure 5 shows that microprobes of the 12.2% v/v PVAc/TNW nanocomposite (without electrical insulation or contact layers), with a cross-sectional area of 6625 μm2, and an angular tip could readily be inserted through the pia mater into the cerebral cortex of a rat, whereas reference probes consisting of neat PVAc buckled under lower loads (∼7 mN) before they could be inserted into the cortical tissue.
Snapshots of a movie that show insertion attempts of microprobes consisting of an adaptive 12.2% v/v poly(vinyl acetate) (PVAc)/tunicate whisker nanocomposite (NC) and (as reference) the neat PVAc polymer into the brain of a rat. (a) Before initial insertion, the NC is about one millimeter above the brain’s surface. (b) Upon movement, the NC indents and then penetrates the pia and cortex. (c) After the completion of movement, the NC is implanted in the brain, and the indentation is relaxed. (d) Before initial insertion, the neat polymer is about one millimeter above the brain’s surface. (e) Upon movement, the neat polymer indents the pia, but quickly buckles. (f) After the completion of movement, the neat polymer is completely buckled. Reprinted with permission from Reference 94. ©2011, IOP Publishing. To review the movie, please visit http://dx.doi.org/10.1557/mrs.2012.97Q1: Please replace this with a doi number. I have placed the mp4 file here: ftp://ftp2.tnq.co.in/CUP/To_TnQ/MRSB/042412/.
We demonstrated the material’s ability to morph while in the rat cortex by implanting samples into the rat cortex for up to 30 minutes, removing them, and quickly placing them in a custom-built microtensile testing setup.Reference Hess, Capadona, Shanmuganathan, Hsu, Rowan, Weder, Tyler and Zorman95 Indeed, such ex vivo samples displayed very similar mechanical morphing profiles as those measured in vitro on a dynamic mechanical analyzer (DMA, Figure 6). Taken collectively, this work demonstrated the first stand-alone material that was capable of facile insertion into the brain, while rapidly softening to more closely match the mechanical properties of its surroundings.
The Young’s modulus of the nanocomposite material (12.2% v/v poly(vinyl acetate)/tunicate whisker) was measured for bulk films with a dynamic mechanical analyzer (DMA) (open squares; bulk materialsReference Shanmuganathan, Capadona, Rowan and Weder73) and microprobes with a micro tensile tester (open circles; explanted microprobes) to reflect the mechanical morphing from stiff to compliant. DMA samples were exposed in situ (i.e., in the DMA setup) to artificial cerebrospinal fluid (ACSF) that was preheated to 37°C,Reference Shanmuganathan, Capadona, Rowan and Weder73 while microprobes were implanted into a rat cortex and explanted for mechanical testing. The x-axis indicates the time of exposure to either ACSF or implantation into the rat cortex, respectively. Reprinted with permission from Reference 94. ©2011, IOP Publishing.
The new materials allowed us to conduct an in vivo study that probed the neural inflammatory response as a function of mechanical characteristics of the probe. To accomplish this, microprobes of the mechanically dynamic 12.2% v/v PVAc/TNW nanocomposite were allowed to remain implanted into the rat cortex for either four or eight weeks (Figure 7).Reference Harris, Capadona, Miller, Healy, Shanmuganathan, Rowan, Weder and Tyler96 The inflammatory and neurodegenerative responses to these materials were compared to contralateral implants of much stiffer (and not adaptive) tungsten microwires of similar dimensions. The latter were coated with a thin layer of neat PVAc to provide matched surface chemistry and roughness. Both implants were tethered to normalize the effects of strain on the tissue, which is relevant for a recording electrode model.
In order to assess the proximity of neurons to the electrode materials that were implanted into rats, we utilized standard biochemical markers for the nucleus of neurons (neuronal nuclei or NeuN) and imaged the tissue sections with a fluorescent microscope. Representative images of NeuN for tissue that had electrode materials implanted for either four (a, b) or eight (c, d) weeks are shown for both nanocomposite (NC) (a, c) and wire (b, d) implants. Each “white” circle represents a neuron. More neurons found close to the hole created by the implant indicates less neuron death at the device/tissue interface. Scale bars are 100 μm. Reprinted with permission from Reference 96. ©2011, IOP Publishing.
Fluorescent immunohistochemistry labeling was used to examine neurons, as well as inflammatory cells. The neuronal nuclei density within 100 μm of the mechanically adaptive device at four weeks post-implantation was significantly larger than that of the stiff wire control. At eight weeks post-implantation, the neuronal nuclei density around the nanocomposite was maintained, but the density around the stiff control recovered to match that of the nanocomposite. While the inflammatory-mediated encapsulation of the compliant nanocomposite was less vigorous than to the stiffer wire, as quantified through detection of relevant cells via immunohistochemistry labeling, the mechanically associated cells and proteins appeared to be the most dominant modulators of the response to the compliant nanocomposite. This initial study did not involve any end points beyond a time frame of eight weeks. However, evidence in other labs, as well as our own, suggests that neurodegeneration may increase from 8 to 16 weeks post-implantation.Reference McConnell, Rees, Levey, Gutekunst, Gross and Bellamkonda12 This is consistent with neurotrauma literature demonstrating a biphasic response of the nervous system immune cells, or microglia, resulting in a secondary event of neurodegeneration.Reference Loane and Byrnes97 While our non-dynamic, stiffer controls follow the trend of scar maturation, the dynamic nanocomposites display unique stabilization across both time points analyzed. Therefore, it appears important to probe additional time points past the eight-week mark.
Summary and future directions
Arguably, the most accepted hypothesis for electrode failure relates to the development of the encapsulating scar at the electrode/tissue interface. A growing belief is that the mechanical mismatch between current electrode materials and the cortical tissue mediates scar formation and subsequent neurodegeneration through a variety of potential mechanisms. This has led to a paradigm shift to the development of new electrodes made from softer polymeric materials and has resulted in a number of recently introduced approaches, such as the use of mechanically adaptive polymer nanocomposites or swellable hydrogels as substrates for the electrode. For example, intracortical implants based on the dynamic nanocomposite materials have been shown to be sufficiently stiff to be easily implanted into the brain and subsequently soften to better match the stiffness of the cortex. Initial histological evaluations suggest that mechanically adaptive intracortical neural prosthetics can more rapidly stabilize neural cell populations at the interface than rigid systems, which bodes well for improving the functionality of intracortical devices. However, there are still many open issues that need to be addressed. First of all, the tissue response to such adaptive materials after longer implantation times than previously studied (up to eight weeks) must be explored. Initial studies have suggested a significant advantage with the softer materials, which fluctuated with the maturation of the foreign body response. This suggests the potential utility of combinatorial approaches that explore both dynamic materials and short-term therapeutic measures.
Additionally, if the results confirm that softer implants are indeed better than more rigid devices, as the initial data suggest, then the question becomes what the optimal mechanical properties for cortical electrodes really are—in other words, how soft do they need to be? And is it the surface that matters or the device as a whole? This leads to the question, “What are the main biological processes that are triggered by the mechanical mismatch?” It appears that the answers to these questions are relevant to the field at large, as approaches other than the one discussed here exist to take advantage of the paradigm “softer is better.” It is noted that so far, little has been done to create actual electrodes on the basis of mechanically adaptive substrates, and it is clear that such devices should be engineered and tested as soon as possible. In order for such devices to be routinely clinically viable, significant efforts to better understand the molecular and cellular events that lead to neurodegeneration at the device-tissue interface are needed to promote the development of materials-based and therapeutic strategies to improve device integration and stabilize neural recordings.
Acknowledgments
This work was supported by grant number R21-NS053798 from the National Institute of Neurological Disorders and Stroke (D.J.T.). Additional support was from the Department of Veterans Affairs Medical Center, specifically grant numbers: C3819C, F4827H, and B6344W (J.R.C.) and NSF grants CBET-0828155 (S.J.R./C.W.) and ECS-0621984 (C.Z.). C.W. gratefully acknowledges financial support from the Swiss National Science Foundation (NRP 62: Smart Materials, Nr. 406240_126046) and the Adolphe Merkle Foundation. Finally, some parts of this work were supported by the Department of Biomedical Engineering at Case Western Reserve University through both laboratory startup funds (J.R.C.), as well as a Medtronic Graduate Fellowship.