Quality assurance of intensity-modulated radiotherapy treatment planning: a dosimetric comparison

Published online by Cambridge University Press:  23 July 2018

Saima Altaf
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan Department of Physics, The Women University, Multan, Pakistan
Khalid Iqbal*
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan Department of Clinical & Radiation Oncology, Shaukat Khanum Cancer Hospital & Research Center Lahore, Pakistan
Muhammad Akram
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan
Saeed A. Buzdar
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan
*
Author for correspondence: Khalid Iqbal, Department of Clinical & Radiation Oncology, Shaukat Khanum Cancer Hospital & Research Center, Lahore, Pakistan. Tel: +92 42359 05000. Fax: +92 42359 45206. E-mail: khalid_phy@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Aim

The purpose of this study was to analyse the comparison of intensity-modulated radiation therapy quality assurance (IMRT QA) using Gafchromic® EBT3 film, Electronic portal imaging device (EPID) and MapCHECK®2.

Background

Pretreatment authentication is the main apprehension in advanced radiation therapy treatment plans such as IMRT.

Materials and methods

A total of 20 patients were planned on Eclipse treatment planning system using 6 and 15 MV separately.

Results

Gamma index of EBT3 film results shows the following average passing rates: 97% for 6 MV and 96·6% for 15 MV using criteria of ±5% of 3 mm, ±3% of 3 mm and ±3% of 2 mm for brain. However, by using ±5% of 3 mm and ±3% of 3 mm criteria, the average passing rates were 95·4% on 6 MV and 95·2% on 15 MV for prostate. For EPID, the results show the average passing rates as 97·8% for 6 MV and 97·2% for 15 MV in for brain. In cases in which ±5% of 3 mm and ±3% of 3 mm were used, the average passing rates were 96·6% for 6 MVand 96·1% for 15 MV for prostate. MapCHECK®2 results show average passing rates of 96·4% for 6 and 96·2% for 15 MV, respectively, for brain using criteria of ±5% of 3 mm, ±3% of 3 mm and ±3% of 2 mm, whereas for ±5% of 3 mm and ±3% of 3 mm the average rates are 95·2% for 6 and 94·7% for 15 MV in prostate.

Conclusions

The EPID results are better than the other methods, and hence EPID can be used effectively for IMRT pretreatment verifications.

Information

Type
Original Article
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 Electronic portal imaging device.

Figure 1

Figure 2 MapCHECK®2 (Sun Nuclear Corporation, Melbourne, FL, USA).

Figure 2

Figure 3 Graphical analysis for passing rates between EBT3, EPID and MapCHECK®2 with treatment planning system-calculated dose distributions for 6 MV brain intensity-modulated radiation therapy plans with Gamma criteria.

Figure 3

Table 1 Statistical analysis for passing rates between EBT3, EPID and MapCHECK®2 and treatment planning system-calculated dose distributions for 6 MV brain intensity-modulated radiation therapy plans with Gamma criteria

Figure 4

Table 2 Statistical analysis for passing rates between EBT3, EPID and MapCHECK®2 and treatment planning system-calculated dose distributions for 15 MV brain intensity-modulated radiation therapy plans with Gamma criteria

Figure 5

Figure 4 Graphical analysis for passing rates between EBT3, EPID and MapCHECK®2 with treatment planning system-calculated dose distributions for 15 MV brain intensity-modulated radiation therapy plans with Gamma criteria.

Figure 6

Figure 5 Graphical analysis for passing rates between EBT3, EPID and MapCHECK®2 with treatment planning system-calculated dose distributions for 6 MV prostate intensity-modulated radiation therapy plans with Gamma criteria.

Figure 7

Table 3 Statistical analysis for passing rates between EBT3, EPID and MapCHECK®2 and treatment planning system-calculated dose distributions for 6 MV prostate intensity-modulated radiation therapy plans with Gamma criteria

Figure 8

Figure 6 Graphical analysis for passing rates between EBT3, EPID and MapCHECK®2 with treatment planning system-calculated dose distributions for 15 MV prostate intensity-modulated radiation therapy plans with Gamma criteria.

Figure 9

Table 4 Statistical analysis for passing rates between EBT3, EPID and MapCHECK®2 and treatment planning system-calculated dose distributions for 15 MV prostate intensity-modulated radiation therapy plans with Gamma criteria