Testing the analytical performance of handheld XRF using marine sediments of IODP Expedition 355

Published online by Cambridge University Press:  04 April 2019

A. Hahn*
Affiliation:
MARUM, University of Bremen, Leobener Str. 8, 28359 Bremen, Germany
M. G. Bowen
Affiliation:
International Ocean Discovery Program, Texas A&M University, College Station, TX 77845, USA
P. D. Clift
Affiliation:
Department of Geology and Geophysics, Louisiana State University, 3838 W Lakeshore Dr, Baton Rouge, LA 70808, USA
D. K. Kulhanek
Affiliation:
International Ocean Discovery Program, Texas A&M University, College Station, TX 77845, USA
M. W. Lyle
Affiliation:
College of Earth, Ocean and Atmospheric Sciences, Oregon State University, Corvallis, OR 97331, USA
*
Rights & Permissions [Opens in a new window]

Abstract

Obtaining geochemical profiles using X-ray fluorescent (XRF) techniques has become a standard procedure in many sediment core studies. The resulting datasets are not only important tools for palaeoclimatic and palaeoceanographic reconstructions, but also for stratigraphic correlation. The International Ocean Discovery Program (IODP) has therefore recently introduced shipboard application of a handheld XRF device, making geochemical data directly available to the science party. In all XRF scanning techniques, the physical properties of wet core halves cause substantial analytical deviations. In order to obtain estimates of element concentrations (e.g. for quantitative analyses of fluxes or mass-balance calculations), a calibration of the scanning data is required. We test whether results from the handheld XRF analysis on discrete samples are suitable for calibrating scanning data. Log-ratios with Ca as a common denominator were calculated. The comparison between the handheld device and conventional measurements show that the latter provide high-quality data describing Al, Si, K, Ca, Ti, Mn, Fe, Zn, Rb and Sr content (R2 compared with conventional measurements: ln(Al/Ca) = 0.99, ln(Si/Ca) = 0.98, ln(K/Ca) = 0.99, ln(Ti/Ca) = 0.99, ln(Mn/Ca) = 0.99, ln(Fe/Ca) = 0.99, ln(Zn/Ca) = 0.99 and ln(Sr/Ca) = 0.99). Our results imply that discrete measurements using the shipboard handheld analyser are suitable for the calibration of XRF scanning data. Our test was performed on downcore sediments from IODP Expedition 355 that display a wide variety of lithologies of both terrestrial and marine origin. The implication is that our findings are valid on a general scale and that shipboard handheld XRF analysis on discrete samples should be used for calibrating XRF scanning data.

Information

Type
Original Article
Copyright
© Cambridge University Press 2019
Figure 0

Fig. 1. Bathymetric map of the Arabian Sea and surrounding landmasses from GeoMapApp (Ryan et al.2009). The red stars indicate expedition 355 Site U1456 and the grey shading the approximate extent of the fan (Kolla & Coumes, 1987). After Pandey et al. (2016).

Figure 1

Fig. 2. Cross-plots for an array of commonly used elemental ratios: log-ratios calculated (a) from onshore XRF scanner data v. conventional XRF and (b) from the handheld scanner v. conventional XRF.

Figure 2

Table 1. Correlation coefficients (R2) of raw XRF scanning data and results from the handheld analyser compared with quantitative results from conventional XRF analysis for an array of commonly used elemental log-ratios. NA – not applicable