Retrospective analysis of portal dosimetry pre-treatment quality assurance of hybrid IMRT breast treatment plans

Published online by Cambridge University Press:  27 February 2020

Meghan Koo
Affiliation:
Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada
Johnson Darko*
Affiliation:
Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada
Ernest Osei
Affiliation:
Department of Physics and Astronomy, University of Waterloo, Waterloo, ON, Canada Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON, Canada Department of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, ON, Canada Department of Systems Design, University of Waterloo, Waterloo, ON, Canada
*
Author for correspondence: Johnson Darko, Department of Medical Physics, Grand River Regional Cancer Centre, Kitchener, ON N2G 1G3, Canada. Tel: 519 749 4300x5793. E-mail: johnson.darko@grhosp.onca
Rights & Permissions [Opens in a new window]

Abstract

Background:

The purpose of this study is to evaluate the effectiveness and sensitivity of the Varian portal dosimetry (PD) system as a quality assurance (QA) tool for breast intensity-modulated radiation therapy (IMRT) treatment plans.

Materials and methods:

Four hundred portal dose images from 200 breast cancer patient IMRT treatment plans were analysed. The images were obtained using Varian PortalVision electronic portal imaging devices (EPIDs) on Varian TrueBeam Linacs. Three patient plans were selected, and the multi-leaf collimator (MLC) positions were randomly altered by a mean of 0·5, 1, 1·5 and 2 mm with a standard deviation of 0·1 mm on 50, 75 and 100% of control points. Using the improved/global gamma calculation algorithm with a low-dose threshold of 10% in the EPID, the change in gamma passing rates for 3%/3 mm, 2%/2 mm and 1%/1 mm criterion was analysed as a function of the introduced error. The changes in the dose distributions of clinical target volume and organ at risk due to MLC positioning errors were also analysed.

Results:

Symmetric and asymmetric breast or chest wall plan fields are different in delivery as well as in the QA. An average gamma passing rate of 99·8 ± 0·5 is presented for 3%/3 mm symmetric plans and 96·9 ± 4·5 is presented for 3%/3 mm asymmetric plans. An average gamma passing rate of 98·4 ± 4·3 is presented for 2%/2 mm symmetric plans and 89·7 ± 9·5 is presented for 2%/2 mm asymmetric plans. A large-induced error in MLC positioning (2·0 mm, 100% of control points) results in an insignificant change in dose that would be delivered to the patient. However, EPID portal dosimetry is sensitive enough to detect even the slightest change in MLC positioning error (0·5 mm, 50% of control points).

Conclusions:

Stricter pre-treatment QA action levels can be established for breast IMRT plans utilising EPID. For improved sensitivity, a multigamma criteria approach is recommended. The PD tool is sensitive enough to detect MLC positioning errors that contribute to even insignificant dose changes.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Bar chart of mean gamma passing rates for breast right (BRER) (n = 50), breast left (BREL) (n = 50), chest wall right (CHWR) (n = 50) and chest wall left (CHWL) (n = 50) treatment plans at 3% dose difference (DD) and 3 mm distance to agreement (DTA). The error bars represent the standard deviation of the corresponding values.

Figure 1

Figure 2. Bar chart of mean gamma passing rates for breast right (BRER) (n = 50), breast left (BREL) (n = 50), chest wall right (CHWR) (n = 50) and chest wall left (CHWL) (n = 50) treatment plans at 2% DD and 2 mm DTA. The error bars represent the standard deviation of the corresponding values.

Figure 2

Figure 3. Bar chart of mean gamma passing rates for breast right (BRER) (n = 50), breast left (BREL) (n = 50), chest wall right (CHWR) (n = 50) and chest wall left (CHWL) (n = 50) treatment plans at 1% DD and 1 mm DTA. The error bars represent the standard deviation of the corresponding values.

Figure 3

Table 1. Mean gamma passing rates for breast right (BRER) (n = 50), breast left (BREL) (n = 50), chest wall right (CHWR) (n = 50) and chest wall left (CHWL) (n = 50) of improved gamma and normalised to the centre of the field treatment plans

Figure 4

Figure 4. Bar chart of mean gamma passing rates for breast symmetric (n = 75), breast asymmetric (n = 25), chest wall symmetric (n = 18) and chest wall asymmetric (n = 82) treatment plans at 3% DD and 3 mm DTA.

Figure 5

Figure 5. Bar chart of mean gamma passing rates for breast symmetric (n = 75), breast asymmetric (n = 25), chest wall symmetric (n = 18) and chest wall asymmetric (n = 82) treatment plans at 2% DD and 2 mm DTA.

Figure 6

Figure 6. Bar chart of mean gamma passing rates for breast symmetric (n = 75), breast asymmetric (n = 25), chest wall symmetric (n = 18) and chest wall asymmetric (n = 82) treatment plans at 1% DD and 1 mm DTA.

Figure 7

Table 2. Mean gamma passing rates for breast symmetric (n = 75), breast asymmetric (n = 25), chest wall symmetric (n = 18) and chest wall asymmetric (n = 82) fields

Figure 8

Table 3. Gamma passing values analysed for the comparison of the predicted plans with the induced errors and their corresponding error-free original predicted plans (PR-PR)

Figure 9

Table 4. Mean gamma passing values analysed for the comparison of the measured plans (with and without error) with the predicted original plan (PR-MS)

Figure 10

Table 5. Mean gamma passing values analysed for the comparison of the measured plans with error to that without error (MS-MS)

Figure 11

Figure 7. Dose volume histogram (DVH) of the clinical target volume (CTV) and organ at risk (OAR) (right lung) of the 13 error induced and the original (error free in red) breast right (BRER) treatment plans.

Figure 12

Figure 8. DVH of the CTV and OARs (left lung and heart) of 13 error induced and the original (error free in red) breast left (BREL) treatment plans.

Figure 13

Figure 9. DVH of the CTV and OAR (right lung) of 13 error induced and the original (error free in red) asymmetric chest wall right (CHWR) treatment plans.