Graphene-based microwave coaxial antenna for microwave ablation: thermal analysis

Published online by Cambridge University Press:  19 October 2020

Burak Uzman
Affiliation:
Department of Electrical and Electronics Engineering, KTO Karatay University, Konya, Turkey
Adem Yilmaz
Affiliation:
Department of Electrical and Electronics Engineering, KTO Karatay University, Konya, Turkey
Hulusi Acikgoz*
Affiliation:
Department of Electrical and Electronics Engineering, KTO Karatay University, Konya, Turkey
Raj Mittra
Affiliation:
Department of Electrical and Computer Engineering, University of Central Florida, Orlando, Florida, USA KAU, Jeddah, Saudi Arabia
*
Author for correspondence: Hulusi Acikgoz, E-mail: hulusi.acikgoz@karatay.edu.tr
Rights & Permissions [Opens in a new window]

Abstract

In this study, the problem of backward heating in microwave ablation technique is examined and an electromagnetic solution based on the use of high impedance graphene material is presented for its mitigation. In this context, a one-atom-thick graphene layer is added on the coaxial double slot antenna. In addition to the electromagnetic behavior, thermal effects caused by the graphene-covered antenna are emphasized. The graphene's conductivity being highly dependent on its chemical potential and the relaxation time, a parametric study is performed to determine a range of tolerances within which the graphene-coated antenna outperform a typical graphene-free antenna. The range of values is found to be 0 < μc < 0.5 eV and τ < 0.4 ps, for the chemical potential and the relaxation time, respectively. The backward heating problem being prevented, the ablation region is ensured to be spherical around the tip of the antenna. Effects of the graphene layer to the heat dissipation in the tissue, the necrotic tissue ratio (damage to the cancerous tissue of the caused by electromagnetic energy), and the treatment time using the coaxial double slot antenna were examined. The results show that the heat dissipation is concentrated around the slots (region of cancerous tissue) and a higher necrotic tissue ratio can be achieved with a graphene-covered double slot antenna in a shorter time.

Information

Type
Antenna Design, Modelling and Measurements
Copyright
Copyright © The Author(s), 2020. Published by Cambridge University Press in association with the European Microwave Association
Figure 0

Fig. 1. Conductivity versus chemical potential for different relaxation time (a), Conductivity versus relaxation time for different chemical potential (b).

Figure 1

Fig. 2. Double slot coaxial antenna: perpendicular cut cross-section (a), parallel cut cross-section with its dimensions (b).

Figure 2

Table 1. Tissue and antenna parameters

Figure 3

Fig. 3. The effect of length (a) and position (b) of graphene on the efficiency of MCA.

Figure 4

Fig. 4. Picture depicting where on the antenna the surface current is computed (a), picture depicting where the SAR is computed in the tissue (b).

Figure 5

Fig. 5. Surface current density along the antenna calculated after the slots and toward the feed line (“normal” refers to the graphene-free antenna) (a) and the SAR all the way along the antenna for different chemical potentials with τ = 0.1 ps (b) and for different relaxation times with μc = 0.5 eV (c).

Figure 6

Table 2. Tissue parameters in penn equation

Figure 7

Fig. 6. Temperature distribution in the tissue along the antenna.

Figure 8

Fig. 7. Pictures depicting where the necrotic tissue rate is computed in the tissue, between the slots (a), after the graphene (b).

Figure 9

Fig. 8. Temperature distribution (a), tissue damaged rate (b), at a 1.5 mm distance, between slots.

Figure 10

Fig. 9. Temperature versus time (a), tissue damaged rate (b), computed at a radial distance of 1.5 mm, after the graphene.

Figure 11

Fig. 10. 2D damage tissue rate: without graphene (a), with 0 eV graphene-coated antenna (b).

Figure 12

Fig. 11. SAR distribution (a), Temperature distribution (b), at 1.5 mm distance from antennas.

Figure 13

Fig. 12. 2D damaged tissue ratio: 0 eV Graphene (a), Choke (b), Sleeve (c).