Early and mid-term outcomes of transcatheter closure of perimembranous ventricular septal defects using the Lifetech™ Konar-MF Occluder device (MFO)

Published online by Cambridge University Press:  16 November 2022

Ender Odemis*
Affiliation:
Faculty of Medicine, Department of Pediatric Cardiology, Koc University, Turkey
Mete Han Kizilkaya
Affiliation:
Faculty of Medicine, Department of Pediatric Cardiology, Koc University, Turkey
*
Author for correspondence: Ender Odemis, Faculty of Medicine, Department of Pediatric Cardiology, Koc University, Turkey. E-mail: odemisender@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Background:

Transcatheter closure of perimembranous ventricular septal defects is challenging. Double disk-designed devices have high adverse event rates; therefore, research into new devices persists. One such devise is the LifetechTM Konar-MF Occluder device (MFO), which is increasingly used. The purpose of this study is to present mid-term results of MFO for transcatheter closure of patients with perimembraneous ventricular septal defect.

Patients and methods:

Records of the 52 patients with perimembraneous ventricular septal defect who had undergone transcatheter closure with MFO were reviewed. Demographic, echocardiographic, and procedure details were investigated. Three years follow-up results were recorded.

Results:

We closed the perimembraneous ventricular septal defect in 51 of 52 patients (98%). The femoral venous approach was used in 27 patients (53 %) whilst no arteriovenous loop was established in the remaining patients. No significant procedure-related complication occurred. On day 1 echocardiography, the residual ventricular septal defect rate was 31%. Mean±SD (range) follow-up duration was 36 ± 7,9 (18–54) months. In the follow-up, at 6th months, only four patients had hemodynamically insignificant residual defects. No severe dysrhythmia was detected including complete heart block. A right bundle branch block pattern was seen in one patient.

Conclusion:

This study showed that MFO is a safe and effective device in the transcatheter treatment of perimembraneous ventricular septal defect’s with mid-term follow-up.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2022. Published by Cambridge University Press
Figure 0

Figure 1. Picture of the MFO Konar device. Screws on the both side of the device allows to use both arterial and venous approach.

Figure 1

Figure 2. pmVSD closure procedure steps with venous (antegrade) approach: (a) Left ventricular injection in left anterior oblique 40 and cranial 20 degrees. (b) Partially exposing the device in the ascending aorta and checking the position of the aortic valve by performing a hand injection (c) lean on the devices’ left disc component to the defect and controlling by left ventricular injection. (d) Exposing of the right ventricular disk and just before releasing (e) control angiogram showing no residual shunt.

Figure 2

Figure 3. pmVSD closure procedure steps with the arterial (retrograde) approach: (a) Left ventricular injection in left anterior oblique 40 and cranial 20 degrees (b) exposing the device discs under transthoracic echocardiographic guidance (c) Control left ventricular angiogram shows no residual shunt.

Figure 3

Table 1. Demographic and echocardiographic features of the patients. Mean ± standart deviation (range).

Figure 4

Table 2. Procedural details of the patients mean ± standart deviation (range).

Figure 5

Table 3. Follow-up results.