Hostname: page-component-74656f9884-c52kr Total loading time: 0 Render date: 2026-11-09T14:15:45.490Z Has data issue: false hasContentIssue false

Temperature compensation of electrical conductivity in glacial meltwaters

Published online by Cambridge University Press:  20 January 2017

C. C. Smart*
Affiliation:
Department of Geography, University of Western Ontario, London, ontario N6A 5C2, Canada
Rights & Permissions [Opens in a new window]

Abstract

Temperature compensation built into many conductivity meters becomes inaccurate in low-temperature waters typical of glacial melt streams. Experiments using simulated glacial waters show a linear dependence of electrical conductivity on temperature from 0.3° to 25°C. The slope of the temperature–conductivity relation is linearly dependent on the conductivity of the solution, allowing a numerical or analytical temperature correction to be made. To minimize error introduced by temperature compensation, measurements of electrical conductivity in glacial streams should be corrected to a low standard temperature, and 0°C is suggested.

Information

Type
Research Article
Copyright
Copyright © International Glaciological Society 1992
Figure 0

Fig. 1. Sample plot of electrical conductivity against temperature for 1.0 mg l−1 sediment in distilled water. Only every tenth data point is marked. Note anomalous points during initial equilibration. These were excluded from the analysis.

Figure 1

Fig. 2. Collective results of all experiments. The lines are described by Equations (1) and summarized in Table 1. Lines represent unfiltered distilled-water mixtures unless marked “Tap” for tap water. Figures give sediment concentration in mg l−1. 1000(F) indicates a 1000mg l−1 distilled-water sample filtered before analysis.

Figure 2

Table 1. Results of electrical conductivity experiments with conventional temperature-correction factor, and intercept (d) and coefficient (b) of best-fit line

Figure 3

Fig. 3. Conventional electronic conductivity correction factors plotted against base conductivity (d = EC0). Key indicates water type, sediment concentration in mg 1−1and filtration (F).

Figure 4

Fig. 4. Thermal coefficient (b) versus base conductivity (d = EC0). ( Key as for Figure 3.)