Hybrid perventricular muscular ventricular septal defect closure using the new multi-functional occluder

Published online by Cambridge University Press:  13 August 2020

Raymond N. Haddad*
Affiliation:
M3C-Necker, Hôpital Universitaire Necker-Enfants malades, Université de Paris, Paris, France
Régis Gaudin
Affiliation:
M3C-Necker, Hôpital Universitaire Necker-Enfants malades, Université de Paris, Paris, France
Damien Bonnet
Affiliation:
M3C-Necker, Hôpital Universitaire Necker-Enfants malades, Université de Paris, Paris, France
Sophie Malekzadeh-Milani
Affiliation:
M3C-Necker, Hôpital Universitaire Necker-Enfants malades, Université de Paris, Paris, France
*
Author for correspondence: Dr. Raymond N. Haddad, MD, Unité médico-chirurgicale de cardiologie congénitale et pédiatrique, centre de référence des malformations cardiaques congénitales complexes – M3C, Hôpital universitaire Necker-Enfants malades, 149, rue de Sèvres, Paris 750015, France. Tel: +33 7 53 15 95 04, +961 70 605 800; Fax: +33 1 44 49 47 30. E-mail: raymondhaddad@live.com
Rights & Permissions [Opens in a new window]

Abstract

The hybrid perventricular approach for the closure of trabecular ventricular septal defects is an attractive treatment modality for small children. Worldwide experience has shown that procedure success is influenced by the defect anatomical accessibility, operators’ expertise, and device technical features. In May 2018, a new promising device, the KONAR-Multi-functional™ ventricular septal defect occluder (Lifetech, Shenzhen, China), obtained CE-marking for septal defect transcatheter closure after the first-in-man implantation in 2013. Herein, this is the first report of successful perventricular closure of ventricular septal defect using this new device in a child with significant co-morbidities.

Information

Type
Brief Report
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1.Ultrasound views. (a) 4-mm residual muscular VSD in a small baby after surgical repair of a TGA complex variant. Note the single left ventricular entry (b) and multiple right ventricular trabeculated exists (c). TGA = transposition of the great arteries; VSD = ventricular septal defect.

Figure 1

Figure 2.Fluoroscopic view. Perventricular access with catheter and guidewire combination passing through the defect into the left ventricular outflow tract.

Figure 2

Figure 3.KONAR-MF™ VSD occluder (Lifetech, Shenzhen, China). (a) A small model of the device (with no PTFE membrane inside) connected to the delivery cable from its right side. (b) Schematic illustration of the device configuration: D, left retention disk diameter; D2, waist diameter (left ventricular side); D1, waist diameter (right ventricular side); L, 4-mm waist length. PTFE = polytetrafluoroethylene; VSD = ventricular septal defect.

Figure 3

Figure 4.(a) TOE imaging with colour-flow Doppler mode showing an unreleased 5/3-mm KONAR-MF™ device in place with no residual shunt. Fluoroscopic view showing stable positioning before (b) and after release (c). TOE = transoesophageal echocardiography.

Figure 4

Figure 5.Ultrasound control showing stable device position 48 hours after deployment.