Apparatus for routine measurements of the thermal conductivity of ice cores

Published online by Cambridge University Press:  14 September 2017

Crescenzo Festa
Affiliation:
Department of Chemistry, University of Pisa, Via Risorgimento, 1-56126 Pisa, Italy
Aristide Rossi
Affiliation:
CNR - International Institute for Geothermal Research, Piazza Solferino 2,1-56126 Pisa, Italy
Rights & Permissions [Opens in a new window]

Abstract

An apparatus is described for measuring the thermal conductivity of ice by the transient hot-wire method. Thermal conductivity A, is determined by tracking the thermal pulse induced in the sample by a heating source consisting of a platinum resistor. A central segment of the same platinum heating resistor acts also as a thermal sensor. A heat pulse transferred to the ice for a period of 40s gives a maximum temperature increment of about 7-14°C. In good experimental conditions, the expected reproducibility of the measurements is within ±3%. The accuracy of the method depends on whether the instrument has been calibrated by reliable standard samples, certified by absolute methods.

Information

Type
Research Article
Copyright
Copyright © International Glaciological Society 1999
Figure 0

Fig. 1.Block diagram of the apparatus for measuring thermal conductivity.

Figure 1

Fig. 2.Measurement of λ at different temperatures with: (1) silica glass (fused quartz) standard samples; (2) Zerodur glass samples. For the silica standard samples the reference λ values are also indicated as: (3) certified by Showa Denko K.K., Tokyo, Japan; (4) recommended by National Standard Reference Data System, U.S. National Bureau of Standards 8 (1966); (5) curve of λ as from Ratcliffe (1959).

Figure 2

Fig. 3.Values of λ measured on two samples of distilled-water ice (a,b), a sample of tap-water ice (c) and a Zirconia reference sample (d).

Figure 3

Fig. 4.Measurement cycle. Typical temperature rise during the 40 s period of heating impulse: (a) temperature measured on the flat wire; (b) temperature at the surface of the sample.

Figure 4

Fig. 5.Scheme of thermal conductivity measurement of an ice-core sample.