Parameters optimization and trajectory planning of a novel 3-UPU parallel mechanism for ankle rehabilitation

Published online by Cambridge University Press:  31 August 2023

Xuechan Chen
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
Yu Guo
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
Chao Xin
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
Chaoyang Ji
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
Jianxun Fan
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
Ziming Chen*
Affiliation:
School of Mechanical Engineering, Yanshan University, Qinhuangdao, Hebei, China. Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao, Hebei, China.
*
Corresponding author: Ziming Chen; Email: chenzm@ysu.edu.cn
Rights & Permissions [Opens in a new window]

Abstract

Ankle rehabilitation robots are widely used due to nerve injuries and sports injuries leading to decreased mobility of the ankle joint. However, the motion of most ankle rehabilitation robots has distinctions with human ankle physiological structure. In order to achieve more accurate rehabilitation training of the ankle joint, this paper proposes a novel 3-UPU parallel rehabilitation mechanism. In a certain range, the mechanism can perform rotation around any axis within the midplane, which means that the mechanism can achieve non-fixed-point rotation around the instantaneous axis of the ankle joint. The mechanism has three degrees of freedom and can perform ankle pronation/supination and inversion/eversion movements. Taking into account the structural differences of different human bodies, the rotating axis of the mechanism can be adjusted in both height and angle. Then, the workspace of the mechanism was solved, and the size parameters of the mechanism are analyzed based on the characteristics of the size parameters of the mechanism and the motion range of the ankle. A genetic algorithm was employed to optimize the mechanism’s parameters. Next, the motion trajectory of the mechanism was planned, and the length change of the mechanism driving pair during the motion planning of the angle was obtained through kinematics simulation. Finally, experimental verification of the above rehabilitation training methods indicates that the mechanism meets the requirements of ankle rehabilitation.

Information

Type
Research Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Figure 1. The rotation axes of the ankle.

Figure 1

Table I. Range of ankle movement.

Figure 2

Figure 2. Structure sketch diagram of the mechanism.

Figure 3

Figure 3. Middle constraint plane M

Figure 4

Figure 4. Schematic diagram of ankle joint movement.

Figure 5

Figure 5. Position adjustment of the mechanism.

Figure 6

Figure 6. A double rocker adjustment mechanism.

Figure 7

Figure 7. Left/right foot rehabilitation mode.

Figure 8

Table II. Body size parameters.

Figure 9

Figure 8. Ankle rehabilitation mechanism.

Figure 10

Figure 9. Relationship between parameter R and rotation angle of mechanism.

Figure 11

Figure 10. The moving platform of the mechanism dimension parameter.

Figure 12

Figure 11. Relationship of mechanism parameters.

Figure 13

Table III. Optimized parameters of mechanism structure.

Figure 14

Figure 12. Parameter optimization results.

Figure 15

Figure 13. Maximum angle of mechanism.

Figure 16

Figure 14. The 3-UPU workspace.

Figure 17

Figure 15. Analysis of planning function.

Figure 18

Table IV. Specific structural parameters of the electric push rode.

Figure 19

Figure 16. Variation of the length of the pronation/supination motion mechanism with time.

Figure 20

Figure 17. Angle change of movable platform during pronation movement.

Figure 21

Figure 18. Rehabilitation training experiment.

Figure 22

Figure 19. Rotation angle of the mechanism.