Total scattering experiments on glass and crystalline materials at the ESRF on the ID11 Beamline

Published online by Cambridge University Press:  22 December 2014

Andrea Bernasconi*
Affiliation:
ESRF – The European Synchrotron – Grenoble, France
Jonathan Wright
Affiliation:
ESRF – The European Synchrotron – Grenoble, France
Nicholas Harker
Affiliation:
ESRF – The European Synchrotron – Grenoble, France
*
a) Author to whom correspondence should be addressed. Electronic mail: andrea.bernasconi@esrf.fr

Abstract

ID11 is a multi-purpose high-energy beamline at the European Synchrotron Radiation Facility (ESRF). Owing to the high-energy X-ray source (up to 140 keV) and flexible, high-precision sample mounting which allows small sample–detector distances to be achieved, experiments such as total scattering in transmission geometry are possible. This permits the exploration of a wide Q range and so provides high real-space resolution. A range of samples (glasses and crystalline powders) have been measured at 78 keV, first putting the detector as close as possible to the sample (~10 cm), and then moving it vertically and laterally with respect to the beam in order to have circular and quarter circle sections of diffraction rings, with consequent Q MAX at the edge of the detector of about 16 and 28 Å−1, respectively. Data were integrated using FIT2D, and then normalized and corrected with PDFgetX3. Results have been compared to see the effects of Q-range and counting statistics on the atomic pair distribution functions of the different samples. A Q of at least 20 Å−1 was essential to have sufficient real-space resolution for both type of samples while statistics appeared more important for glass samples rather than for crystalline samples.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2014 
Figure 0

Table I. Molar composition (expressed as oxide fractions) of the two investigated glass samples.

Figure 1

Figure 1. (Colour online) Examples of 2D images collected with the beam in the CCD corner on sample A (left side) and with the beam in the CCD center on ZnO sample (right side).

Figure 2

Figure 2. (Colour online) Sample A raw data without (red line) and with (black line) the application of incident angle correction using Eq. (4). Effects of phosphor and faceplate corrective terms of Eq. (4) are also plotted as dashed and dotted lines, respectively.

Figure 3

Figure 3. (Colour online) Instrumental resolution obtained by Pseudo-Voigt fit (used in 0–8 Å−1 range) and Gaussian fit (used in 8–13 Å−1 range) of Si NIST SRM640c peaks at different Q-values (beam was in the CCD corner). For each selected peak, the R2 of the Pseudo-Voigt/Gaussian fit to determine FWHM was always above the values of 0.99. Afterwards, 20 different points have been fitted with a 2nd degree polynomial function (red line).

Figure 4

Table II. Integration strategy adopted for evaluate QMAX role (upper part) and statistic role (lower part) during processing with FIT2D.

Figure 5

Figure 4. (Colour online) Zincite G(r)s without (red line) and with (black line) the application of incident angle correction using Eq. (4). Curves have been obtained with a QMAX of 28 Å−1 and an azimuthal range of 90°.

Figure 6

Figure 5. (Colour online) Total scattering F(Q) function of nanocrystalline CeO2, using an azimuthal integration range of 90° and QMAX of 28 Å−1, on the left side. Comparison of G(r) resulting from Fourier transform of the same F(Q) but changing the QMAX from 28 Å−1 (black line), 24 Å−1 (red line), and 20 Å−1 (blue line).

Figure 7

Table III. Full-width half-maximum (FWHM) of the first metal–metal peak in G(r) and refined metal and oxygen UISO parameter in nanocrystalline CeO2 and industrial grade ZnO, as function of QMAX and azimuthal range. Moreover, the RW values from the PDFGUI fit are summarized. Fits have been performed in the 0.5–30 Å range.

Figure 8

Figure 6. G(r) magnification if QMAX of 20 Å−1 (black line) and 16 Å−1 (dashed line) are used. Owing to the different real-space resolution, the former is able to show both Ce–Ce and Ce–O peaks, located at about 9.35 and 9.65 Å, respectively, whereas the latter is able to show only the Ce–Ce peak.

Figure 9

Figure 7. (Colour online) Samples A and B G(r)s as function of different adopted QMAX, obtained with PDFGetX3 program. Black curve is obtained with QMAX of 16 Å−1, putting the beam in the CCD center, while blue and red curves have been obtained by putting the beam in the CCD corner, using an azimuthal range of 90° and a QMAX of 28 (blue line) and 20 Å−1 (red line).

Figure 10

Figure 8. (Colour online) On the left side, the different noise in high Q in S(Q) is displayed, whereas on the right, differences in the resultant G(r)s are shown in the long range. The black line refers to an azimuthal range of 90°, averaging 20 images the red line refers to a reduced azimuthal range of 20°, averaging 20 images; whereas the blue line refers to an azimuthal range of 90°, considering one single image.