Enhancement of a low-fidelity surgical simulator. Is it possible?

Published online by Cambridge University Press:  27 January 2021

L M Evans*
Affiliation:
ENT Department, University Hospital of Wales, Cardiff, UK
D Owens
Affiliation:
ENT Department, University Hospital of Wales, Cardiff, UK
*
Author for correspondence: Dr Louise M Evans, ENT Department, University Hospital of Wales, Heath Park Way, Cardiff CF14 4XW, UK E-mail: louiseevans_5@hotmail.co.uk
Rights & Permissions [Opens in a new window]

Abstract

Background

Simulation training has become a core component in the training of ENT surgeons. It provides the opportunity for the repetitive practice of a surgical technique. Simulators are broadly categorised into low- and high-fidelity simulators. A method using a home microprocessor to enhance a low-fidelity surgical simulator is introduced.

Method

The Yorick tonsil tie trainer was enhanced using an Arduino microcontroller attached to the simulated inferior pole of the tonsil. The Arduino was coded to give a visual stimulus when linear motion exceeded parameters. The prototype simulator was tested to gain information on whether the enhancement could identify differences between novice and expert users.

Conclusion

An enhanced low-fidelity tonsil trainer was produced using a low-cost, simple home microprocessing board. The enhanced simulator gives objective feedback allowing for self-directed learning. Further research is required to evaluate the benefits of these enhancements above non-enhanced simulation training.

Information

Type
Short Communications
Copyright
Copyright © The Author(s), 2021. Published by Cambridge University Press
Figure 0

Fig. 1. The Yorick tonsil tie trainer was enhanced using an Arduino microcontroller/microprocessor board. The Arduino microprocessor was coded using Scratch programming language to assess for linear acceleration in the X, Y and Z axes.

Figure 1

Fig. 2. The microcontroller board was attached to an accelerometer sensor that was attached to the simulated inferior tonsil pole (the paper clip represents the inferior tonsil pole).

Figure 2

Fig. 3. When excessive acceleration was reached with the potential to displace the inferior tonsil pole tie, the Arduino microprocessor switched on a light-emitting diode that was visible to the trainee within the simulation field.

Figure 3

Fig. 4. The light-emitting diode (LED) provided objective feedback: when the LED was on, linear acceleration was too high and risked tie avulsion, and when it was off, it indicated that movement was within acceptable limits.