A new continuous high-resolution detection system for sulphate in ice cores

Published online by Cambridge University Press:  14 September 2017

Matthias Bigler
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Anders Svensson
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Jørgen Peder Steffensen
Affiliation:
Ice and Climate Research, Niels Bohr Institute, University of Copenhagen, Juliane Maries Vej 30, DK-2100 København Ø, Denmark E-mail: bigler@gfy.ku.dk
Patrik Kaufmann
Affiliation:
Climate and Environmental Physics, Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Rights & Permissions [Opens in a new window]

Abstract

Sulphate (SO42–) is a major ion found in polar ice cores and is related to different aerosol sources and processes. Explosive volcanic eruptions, even far away, can cause distinct sulphate peaks in ice core records. Thus, a robust sulphate detection system which is suitable for fieldwork and which enables the measurement of sulphate at high temporal resolution is of great interest. In this study, we present the adaptation of a new continuous flow analysis system for sulphate that is based on a spectrophotometric method using dimethylsulfonazo III and an inline reactor column containing barium sulphate particles. The method shows a detection limit of ∽70 ng g–1 and a linear range up to at least 3000 ng g–1. It is simple, robust and less prone to interferences compared to the previously used method.

Information

Type
Research Article
Copyright
Copyright © The Author(s) [year] 2007 
Figure 0

Fig. 1.Setup of the CFA SO42– detection system. Features include the inline Cation Exchange Column (CEC), the overflow debubbler and the inline BaSO4 reactor column. Flow rates are given for each pump tube of the peristaltic pump. The broken line indicates the fibre optic cables from the flow cell to the lamp and to the detector, respectively.

Figure 1

Fig. 2.(a) Spectrometer signal S (number of counts) of a SO42– standard series (dark grey) with given concentration values. The baseline (light grey) is calculated by linearly interpolating between two sections of blank measurements (black) before and after the standards. (b) Calculated absorption values A of the standard series (dark grey) with respect to the zero line (light grey). Plateau values used to calculate the calibration curve (Fig. 3) are highlighted in black. The insertion shows the measured absorption spectrum A (in arbitrary units) with a maximum around the wavelength λ = 659 nm.

Figure 2

Fig. 3.Calibration curve between measured absorption values and given standard concentrations (points). To calculate the linear fit (A = –0.000 103 C) the two highest values were excluded to account for the deviation from linearity at high concentration levels. The insertion contains an enlargement of the lowest concentration values showing a satisfactory linearity (error bars correspond to the standard deviations of the standard plateaus in Fig. 2).