Complete transcatheter closure of a patent arterial duct with subsequent haemolysis

Published online by Cambridge University Press:  25 March 2010

Neven Cace*
Affiliation:
Department of Pediatrics, University Hospital Center Rijeka, Rijeka, Croatia
Vladimir Ahel
Affiliation:
Department of Pediatrics, University Hospital Center Rijeka, Rijeka, Croatia
Iva Bilic
Affiliation:
Department of Gynecology and Obstetrics, University Hospital Center Rijeka, Rijeka, Croatia
*
Correspondence to: Dr N. Cace, MD, Istarska 43, 51000 Rijeka, Croatia. Tel: +385 51 659 111; Fax: +385 51 623 126; E-mail: ncace1@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

We report a development of severe haemolysis after complete transcatheter closure of patent arterial duct. Aortography and echocardiography revealed no signs of residual shunt. Haemolysis occurred a day after the implantation. Aortography was performed and the extrusion of coil in aorta was evident. The extruded part of the coil was surgically removed. No signs of haemolysis remained.

Information

Type
Brief Reports
Copyright
Copyright © Cambridge University Press 2010
Figure 0

Figure 1 Conical duct shown on aortography.

Figure 1

Figure 2 Complete closure of the patent arterial duct 10 minutes after the procedure.

Figure 2

Figure 3 Extrusion of coil loop in aorta with no opacification in the pulmonary artery.

Figure 3

Figure 4 Part of the coil removed from the aorta.