Penumbra width determination of single beam and 201 beams of Gamma Knife machine model 4C using Monte Carlo simulation

Published online by Cambridge University Press:  11 September 2018

Atefeh Mahmoudi
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Islamic Republic of Iran
Alireza Shirazi*
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Islamic Republic of Iran
Ghazale Geraily*
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Islamic Republic of Iran
Tahereh Hadisi nia
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Islamic Republic of Iran
Masoume Bakhshi
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Islamic Republic of Iran
Maryam Maleki
Affiliation:
Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Isfahan University of Medical Sciences, Islamic Republic of Iran
*
Author for correspondence: Ghazale Geraily, Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Iran. Email: gh-geraily@sina.tums.ac.ir
Author for correspondence: Ghazale Geraily, Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Iran. Email: gh-geraily@sina.tums.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

Background

One of the stereotactic radiosurgery techniques is Gamma Knife radiosurgery, in which intracranial lesions that are inaccessible or inappropriate for surgery are treated using 201 cobalt-60 sources in one treatment session. In this conformal technique, the penumbra width, which results in out-of-field dose in tumour-adjacent normal tissues should be determined accurately. The aim of this study is to calculate the penumbra widths of single and 201 beams for different collimator sizes of Gamma Knife machine model 4C using EGSnrc/BEAMnrc Monte Carlo simulation code and comparison the results with EBT3 film dosimetry data.

Methods and materials

In this study, simulation of Gamma Knife machine model 4C was performed based on the Monte Carlo codes of EGSnrc/BEAMnrc. To investigate the physical penumbra width (80−20%), the single beam and 201 beams profiles were obtained using EGSnrc/DOSXYZnrc code and EBT3 films located at isocentre point in a spherical Plexiglas head phantom.

Results

Based on the results, the single beam penumbra widths obtained from simulation data for 4, 8, 14 and 18 mm collimator sizes along X axis were 0·75, 0·77, 0·90 and 0·92 mm, respectively. The data for 201 beams obtained from simulation were 2·61, 4·80, 7·92 and 9·81 mm along X axis and 1·31, 1·60, 1·91 and 2·14 mm along Z axis and from film dosimetry were 3·21, 4·90, 8·00 and 10·61 mm along X axis and 1·22, 1·69, 2·01 and 2·25 mm along Z axis, respectively.

Conclusion

The differences between measured and simulated penumbra widths are in an acceptable range. However, for more precise measurement in the penumbra region in which dose gradient is high, Monte Carlo simulation is recommended.

Information

Type
Original Article
Copyright
© Cambridge University Press 2018 
Figure 0

Figure 1 The piece of EBT3 film at the centre of the special cassette (left panel). The spherical head phantom with the cassette inside located in the 18 mm helmet (right panel).

Figure 1

Figure 2 The view of the fixed and moving (helmet) collimators for 18mm collimator size.3

Figure 2

Figure 3 Simulation of single channel of Gamma Knife unit.

Figure 3

Table 1 Inner and outer aperture size of moving collimators

Figure 4

Figure 4 The view of the 201 sources arrangement of Gamma Knife unit.4

Figure 5

Figure 5 Comparison of measured and simulated profiles for four field sizes along X axis (left panel) and Z axis (right panel).

Figure 6

Table 2 Measured and Monte Carlo (MC) calculated physical penumbra widths (80–20%) of 201 beams along X axis

Figure 7

Table 3 Measured and Monte Carlo (MC) calculated physical penumbra widths (80–20%) of 201 beams along Z-axis

Figure 8

Figure 6 Simulated single beam profiles for 4, 8, 14 and 18-mm collimator sizes along X axis at isocentre depth.

Figure 9

Table 4 Comparison of single and 201 beam physical penumbra widths (80–20%) using Monte Carlo simulation along X axis