Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-05T14:45:45.393Z Has data issue: false hasContentIssue false

Experimental determination of set-up displacements in anthropomorphic phantom in single-isocentre radiosurgery for multiple brain metastases by off-axis Winston–Lutz test: ExacTracTM v.6 versus DynamicTM

Published online by Cambridge University Press:  24 July 2023

J. A. Rojas-López*
Affiliation:
Universidad Nacional de Córdoba, Facultad de Matemática, Física, Astronomía y Computación, Córdoba, Argentina Hospital Almater, Mexicali, Baja California, Mexico
M. A. Chesta
Affiliation:
Universidad Nacional de Córdoba, Facultad de Matemática, Física, Astronomía y Computación, Córdoba, Argentina
C. D. Venencia
Affiliation:
Instituto Zunino, Córdoba, Argentina
*
Corresponding author: J. A. Rojas-López; Email: alejandro.rojas@almater.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

We compare the accuracy of the off-axis Winston–Lutz (WL) test in two versions of ExacTracTM: version 6.0 (ETv6) and Dynamic (ETD) in the same linac (TrueBeam STx®).

Materials and methods:

An upgraded of the ExacTracTM system was done in our institution. It was designed as an off-axis WL test before the update for comparison purposes. A head 3D-printed phantom based on a patient’s computed tomography images was used. Nine metallic fiducials were inserted and distributed on the phantom. Each target (fiducial) was designed an off-axis WL test with eight different gantry/collimator/table combinations. The phantom was placed using two different ETv6 and ETD in the same linac, and cone-beam computed tomography and electronic portal imaging device (EPID) images were acquired. The 2D deviation between the centre of the fiducial and the radiation field was found and compared with the original digital reconstructed radiography (DRR) by the profiles.

Results:

The phantom allows the definition of a procedure to determine off-axis deviations in radiosurgery treatments. The displacements calculated from the WL test showed acceptable values for both versions taking into account 3D displacement tolerances of 1 mm. These values were reached with rigorous quality assurance (QA) linac tests performed routinely that include mechanical, MV/kV and image-guided radiotherapy (IGRT) tests. However, ETD indicated more accurate values for all the targets no matter the distance to the isocentre (3D displacements < 0·5 mm).

Conclusion:

In terms of the IGRT correction without set-up displacements, ETD is up to twice as accurate as the ETv6, showing 3D displacements up to 0·5 mm in all targets.

Information

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Figure 1. Anthropomorphic phantom for off-axis WL test. (a) Plastic holders and fiducials. (b) Phantom interior and metallic fiducials and their plastic spherical holders. (c) Frontal view of the phantom with thermoplastic mask. (d) Lateral view. (e) Frontal view. (f) Axial CT image of the phantom.

Figure 1

Table 1. Gantry, couch and collimator angle combinations at which images of the targets were obtained

Figure 2

Figure 2. Off-axis WL test. (a) Beam eye view (BEV) for the combination collimator 0º, gantry 0º and couch 0º with an MLC square field of 2·6 × 2·6 cm2 and jaw field of 3 × 3 cm2 and (b) BEV for the combination collimator 90º, gantry 90º and couch 0º. (c) Offline review of the image acquired and the DRR. The isocentre is defined in the intersection point of the yellow orthogonal axis. (d) Horizontal profile extracted from (c).

Figure 3

Figure 3. Seven months follow-up of the 3D displacements obtained by WL test at isocentre with the ExacTrac™ v6.0 (red circles) and ExacTrac Dynamic™ (blue triangles).

Figure 4

Figure 4. Comparison of the shifts in both ExacTrac™ (v6.0 and Dynamic) obtained by CBCT and EPID by the off-axis WL test for the nine targets.

Figure 5

Figure 5. 3D displacements obtained for both ExacTrac™ (ETv6 and ETD) obtained by CBCT and EPID. Horizontal lines represent the maximum tolerance allowed (1 mm) and IGRT for SRS intrafraction tolerance (0·5 mm).

Figure 6

Table 2. Mean displacements by applying rotations in both IGRT systems using EPID

Supplementary material: File

Rojas-López et al. supplementary material

Rojas-López et al. supplementary material

Download Rojas-López et al. supplementary material(File)
File 29.5 KB