Chlorine Dioxide Versus Automated Vaporized Hydrogen Peroxide for High-Level Disinfection of Ultrasound Probes in Outpatient Practice

Published online by Cambridge University Press:  23 June 2026

Ivis Call Rosado
Affiliation:
Memorial Healthcare System
Rachel Guran
Affiliation:
Memorial Healthcare System

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

Background: High-level disinfection (HLD) of ultrasound probes is a required patient safety and regulatory practice. Automated vaporized hydrogen peroxide (VHP) systems, widely implemented for HLD since 2011, involve sizable capital costs, substantial maintenance requirements, and lengthy probe turnaround times. Chlorine dioxide foam has been used for ultrasound probe HLD in Europe since the late 1990s and received U.S. Food and Drug Administration clearance in 2023, consistent with the ANSI/AAMI ST58 standard. Evidence describing operational impact and feasibility in U.S. outpatient practice remains limited. Objective: To compare workflow and probe turnaround times between two ultrasound probe HLD methods and to evaluate the feasibility of chlorine dioxide foam implementation, including staff training requirements and adherence to reprocessing protocols. Methods: From October 2025 to January 2026, we conducted a quality improvement study examining the transition from automated VHP to manual chlorine dioxide foam for transducer HLD in an outpatient urology practice. A total of six registered nurses and four medical assistants participated in the study. Training provided to staff included manufacturer-provided online training, 1:1 in-person instruction, and direct observation to confirm competence with the new reprocessing protocol. Workflow steps for both HLD methods were mapped and compared. Outcomes measured included probe turnaround time (defined as the interval from immediate post-use handling to readiness for subsequent clinical use); number of workflow steps; feasibility of protocol-specific staff training; and adherence to the reprocessing protocol. Results: Each of the methods had a comparable number of workflow steps. Average probe turnaround time for chlorine dioxide was four minutes, while the automated VHP system required 14–15 minutes. Most participants completed manufacturer-provided online training. All participants received 15 minutes of 1:1 in-person instruction and direct observation. Overall, direct observation demonstrated high adherence to the reprocessing protocol. The manual foam process was integrated into routine clinical workflows without disruption to patient care. Conclusion: In a single outpatient urology practice, manual chlorine dioxide foam disinfection was feasible to implement and integrated into existing workflows with considerably shorter probe turnaround time while maintaining staff adherence. Results informed our policy on workflow steps, training, and documentation. Observations were conducted over a limited period, and findings may not be generalizable to other clinical settings.

Information

Type
Disinfection/Sterilization
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited.
Copyright
© The Author(s), 2026. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America