Percutaneous closure of ventricular septal defect using LifeTechTM Konar-MF VSD Occluder: initial and short-term multi-institutional results

Published online by Cambridge University Press:  28 July 2021

Masood Sadiq*
Affiliation:
Department of Pediatric Cardiology, The Children’s Hospital and Institute of Child Health, Lahore, Pakistan
Ahmad Usaid Qureshi
Affiliation:
Department of Pediatric Cardiology, The Children’s Hospital and Institute of Child Health, Lahore, Pakistan
Muhammad Younas
Affiliation:
Department of Pediatric Cardiology, CPE Institute of Cardiology, Multan, Pakistan
Sohail Arshad
Affiliation:
Department of Pediatric Cardiology, The Children’s Hospital and Institute of Child Health, Multan, Pakistan
Syed Najam Hyder
Affiliation:
Department of Pediatric Cardiology, The Children’s Hospital and Institute of Child Health, Lahore, Pakistan
*
Author for correspondence: Professor Masood Sadiq MRCP (UK), FRCP(Ed), FRCPCH Department of Paediatric Cardiology, The Children’s Hospital and Institute of Child Health, Ferozepur Road, Lahore, Pakistan. Tel: 0092-42-99231723; Fax: 0092-42-99230358. E-mail: drmasoodsadiq@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

Background:

Transcatheter ventricular septal defect closure remains a complex procedure with potential complications like complete heart block and aortic regurgitation. The ideal device design for such intervention is still evolving.

Aim:

To assess the safety, efficacy, and short-term outcome of ventricular septal defect closure using LifeTechTM multifunctional (KONAR-MFTM) VSD Occluder.

Patients and methods:

In a multicenre study, 44 patients with haemodynamically significant, restrictive ventricular septal defects underwent closure with the KONAR-MFTM device from April, 2019 to March, 2020. Clinical, echocardiographic, and angiographic data were collected and reviewed. Patients were followed up at 1, 3, 6, and 12 months.

Results:

The median age and weight were 8 (1.7–36) years and 20 (11–79) kg. Of 44 patients, 8 (18%) had a high muscular and 36 (82%) had a perimembranous defect, of which 6 had mild prolapse of the right coronary cusp. The median ventricular septal defect size was 8.8 (3.9–13.4) mm. A retrograde approach was adopted in 39 (88.6%) patients. Nine patients (20.5%) had a small residual leak and there was a slight increase in aortic regurgitation in one patient. One device, which embolised to pulmonary artery was retrieved, and the defect was closed with a larger device. At a median follow-up of 13 (5–18) months, the residual leak persisted in 1 (2.3%) patient. Mild aortic regurgitation in one patient remained unchanged. There were no major complications.

Conclusion:

Percutaneous closure of ventricular septal defect using KONAR-MFTM device is safe and effective in short and midterm follow-up including selected patients with perimembranous defect and mild prolapse of the right coronary cusp.

Information

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press
Figure 0

Figure 1. Device closure of a perimembranous defect using retrograde approach. (a) Left ventriculogram showing membranous septal defect. (b) Delivery sheath passed across VSD into RV. (c) RV end of device being released in RV, while LV end still attached to delivery cable within delivery sheath. (d) Final position of device after deployment completely occluding the defect.

Figure 1

Figure 2. Device closure of a high muscular defect using retrograde approach. (a) Left ventriculogram showing high muscular VSD. (b) LV outflow angiogram showing high muscular VSD. (c) Crossed through retrograde approach with glide wire and engaged in main pulmonary artery. (d) Device's RV disc released in RV. (e) LV disc released in LV. (f) Final position checked prior to release. (g) Final result confirmed on LAO 45° projection ventriculogram. (h) Final result confirmed on LAO 60° projection ventriculogram.

Figure 2

Table 1. Demographic characteristics and pre-procedural details of patients.

Figure 3

Table 2. Procedural details.

Figure 4

Table 3. Complications.

Figure 5

Figure 3. Gradual normalisation of LV diastolic dimensions in 18 months’ follow-up.

Figure 6

Table 4. Follow-up data.

Figure 7

Figure 4. Echocardiographic picture of pmVSD with RCC prolapse at follow- up. (a) Parasternal long axis 2D. (b) Parasternal long axis with color Doppler. (c) Parasternal short axis 2D. (d) Parasternal short axis with color Doppler. (e) Parasternal short axis with 3D showing device in situ, no residual defect and no aortic insufficiency.