Percutaneous self-expandable valve implantation in patients with stented pulmonary arteries: case examples

Published online by Cambridge University Press:  17 January 2025

Judyta Szeliga
Affiliation:
Department of Paediatric Cardiology, University Children’s Hospital, Jagiellonian University Medical College, Krakow, Poland
Marc Gewillig
Affiliation:
Department Pediatric Cardiology, University Hospitals Leuven, Leuven, Belgium
Sebastian Góreczny*
Affiliation:
Department of Paediatric Cardiology, University Children’s Hospital, Jagiellonian University Medical College, Krakow, Poland
*
Corresponding author: Sebastian Góreczny; Email: sebastian.goreczny@uj.edu.pl
Rights & Permissions [Opens in a new window]

Abstract

CHDs affecting the right ventricular outflow tract often require repeated surgical or transcatheter interventions due to pulmonary insufficiency or stenosis. This study presents percutaneous implantation of large self-expanding valves in patients with complex right ventricular outflow tract anatomy after prior pulmonary branch stenting.

Information

Type
Brief Report
Copyright
© The Author(s), 2025. Published by Cambridge University Press
Figure 0

Figure 1. Percutaneous implantation of a self-expanding valve in patient with tetralogy of Fallot after surgical patch repair and stent implantation in the left pulmonary artery (LPA). (a) An initial angiography showed unobstructed pulmonary arteries with a wide stent (white arrowheads) in the LPA, short pulmonary trunk (black arrowheads) and significant retrograde contrast flow (asterisks) to the enlarged right ventricle. (b) Through a 26 Fr DrySeal sheath (Gore, black arrow) placed at the level of the native pulmonary valve, a 36 – 25 mm Venus P-Valve (MedTech) was introduced to the proximal right pulmonary artery. Opening of the distal flare was initiated at the level of the top edge of the stent in the LPA (white arrow). (c) With continuous pushing on the system the entire distal flare was gradually exposed to seat into the stent in the LPA. (d) After the entire valve was exposed the distal flare (black arrow) fit well in the stent. (e) Control angiography from the MPA showed unobstructed contrast flow to both pulmonary arteries with a trace of central regurgitation. (f) Control angiography from the right ventricle confirmed perfect position of the valve with even contrast flow to both pulmonary arteries.

Figure 1

Figure 2. Percutaneous implantation of a self-expanding valve in patient with tetralogy of Fallot after surgical patch repair and stent implantation in the right pulmonary artery (RPA). (a, b) The proximal end of the stent in the RPA was adapted with a sizing balloon to create a large landing space for the distal flare of the Venus-P Valve (MedTech). (c) A 32-30 mm valve was introduced through a 24 Fr DrySeal sheath (Gore) to the proximal left pulmonary artery. Opening of the distal flare was initiated at the level of the top edge of the stent in the RPA (white arrow). (d) Exposing the remainder of the valve was performed in such a way as to keep the distal markers of the valve at the lower edge of the stent. (e) After complete exposure of the valve the distal markers (white arrows) matched the lower end of the stent and the distal flare fit in the RPA stent. (f) Control angiography from the MPA showed unobstructed contrast flow to both pulmonary arteries with a trace of central regurgitation.

Supplementary material: File

Szeliga et al. supplementary material 1

Szeliga et al. supplementary material
Download Szeliga et al. supplementary material 1(File)
File 16.7 MB
Supplementary material: File

Szeliga et al. supplementary material 2

Szeliga et al. supplementary material
Download Szeliga et al. supplementary material 2(File)
File 25.1 MB