Detection of biological abnormalities using a near-field microwave microscope

Published online by Cambridge University Press:  21 May 2018

Fatemeh Kazemi
Affiliation:
Faculty of Electrical and Computer Engineering, University of Sistan and Baluchestan, Zahedan, Iran
Farahnaz Mohanna*
Affiliation:
Faculty of Electrical and Computer Engineering, University of Sistan and Baluchestan, Zahedan, Iran
Javad Ahmadi-Shokouh
Affiliation:
Faculty of Electrical and Computer Engineering, University of Sistan and Baluchestan, Zahedan, Iran
*
Author for correspondence: F. Mohanna, E-mail: F_mohanna@ece.usb.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

Microwave and millimeter-wave reflectometry is one of the potential techniques for the diagnosis and detection of biological abnormalities, such as subcutaneous masses or cancerous tumors in human body. In this paper, a high-quality microwave sensor based on planar misrostrip resonator is designed, fabricated, and successfully tested with different kinds of biological samples. The proposed sensor has unique properties such as small size, simple fabrication, non-contact with a sample, excellent de-coupling from surroundings, and high microwave power is directly coupled into the tissue. Two experiments are performed for the detection of visible and hidden “lipoma” (fat masses), and minimum size of “lipoma” that is diagnosed by the sensor is obtained as well. In this regard, a two-dimensional image of hidden “lipomas” with amplitude contrast about 30 dB and frequency shifts contrast about 60 MHz at a λ/10 (at 13.5 GHz) stand-off distance is provided. Finally, a measurement scenario for the detection of skin cancer based on the artificial model using different layers of raw chicken with different water content is described. Results show that the proposed microscope is easy to fabricate and provide a low-cost solution for fast and accurate skin cancer detection.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2018 
Figure 0

Fig. 1. Geometry of a microwave sensor.

Figure 1

Table 1. Dimensions (mm) of the sensor

Figure 2

Fig. 2. Rectangular near-field pattern of the sensor at 13.5 GHz (H-plane).

Figure 3

Fig. 3. Reflection coefficient of the sensor and photograph of the manufactured sensor.

Figure 4

Fig. 4. (a) Resonant frequency and amplitude of the reflection coefficient is changed when the sensor is placed in front of the MUT. (b) Interaction between near-field sensor and MUT.

Figure 5

Fig. 5. (a) Series of lumped RLC models of a microwave sensor. (b) Circuit model in the presence of a dielectric MUT.

Figure 6

Fig. 6. (a) Reflection coefficient of the sensor in the presence of different biological tissues at 2 mm stand-off distance. (b) E-field amplitude at 2 mm stand-off distance in front of fat mass along the muscle tissue.

Figure 7

Fig. 7. (a) Schematic representation and (b) photograph of our experimental set-up.

Figure 8

Fig. 8. (a) Photographs of “lipomas” with different sizes. (b) Magnitude of reflection coefficient and frequency shift (Δf).

Figure 9

Fig. 9. (a) Photograph of “lipomas” with different sizes below skin layer. (b) Magnitude of reflection coefficient and frequency shift (Δf).

Figure 10

Fig. 10. Photograph of “lipomas” with different sizes, (a) without cover, (b) with cover by a layer of skin, and a microwave image from, (c) measurement of the S11 magnitude, (d) measurement of the frequency shift (Δf).

Figure 11

Fig. 11. Effect of water content in the skin on the magnitude and resonant frequency of reflection coefficient.

Figure 12

Fig. 12. (a) Artificial model to detect skin cancer at the early stage. (b) Magnitude of S11 and frequency shift (Δf).