Simplified sphericity index of PTV can be a predictor for the gradient index in SRS/SRT treatment of brain metastases

Published online by Cambridge University Press:  18 November 2024

Bijay Kumar Barik*
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
Anindya Kumar Baral
Affiliation:
M.K.C.G. Medical College and Hospital, Berhampur, Odisha, India
Santosh Kumar Parida
Affiliation:
Department of Physics, ITER, Siksha O Anusandhan Deemed to be University, Bhubaneswar, Odisha, India
Dillip Kumar Sahoo
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
Sandip Kumar Barik
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
Deepak Kumar Das
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
Saroj Kumar Das Majumdar
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
Dillip Kumar Parida
Affiliation:
All India Institute of Medical Sciences, Bhubaneswar, Odisha, India
*
Corresponding author: Bijay Kumar Barik; Email: medphy_bijay@aiimsbhubaneswar.edu.in
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

To study the simplified sphericity index (SSI) of planning target volume (PTV) and correlate it with the gradient index (GI) for stereotactic radiosurgery (SRS)/stereotactic radiotherapy (SRT) treatment of brain metastasis.

Materials & Methods:

A collection of fifteen brain metastasis cases previously treated with SRS/SRT by volumetric-modulated arc therapy (VMAT) technique was included in the analysis. All the previous plan data from Monaco 6.2.1.0 TPS were used for re-planning and computation of SSI and GI. Pearson’s correlation analysis was performed by using OriginPro 8.5 software, and the outcomes were tabulated.

Results:

The statistical analysis and linear fitting of data show a negative linear correlation between SSI and GI, taking SSI as the independent variable and GI as the dependent variable. Pearson’s correlation coefficient (r) was found to be -0.91563 with a p-value of 0.0000124 showing strong statistical significance.

Conclusion:

It is observed that the GI of the PTV improves as the SSI increases, that is, when the target volume approaches a perfect sphere. Calculating the SSI of the target before planning may help in predicting the GI which may guide making crucial decisions regarding PTV dose prescription and acceptance criteria for organs-at-risk dose tolerance.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1.Measurement of PTV dimensions in three planes: (a) the axial (b) the sagittal, and (c) the coronal section.

Figure 1

Table 1.Statistics of dataset

Figure 2

Figure 2.Plot of SSI versus GI showing a negative linear relationship.