Hostname: page-component-74656f9884-rdgjk Total loading time: 0 Render date: 2026-11-08T05:55:41.958Z Has data issue: false hasContentIssue false

Review on the feasibility of using PRESAGE® dosimeter in various radiotherapy techniques

Published online by Cambridge University Press:  16 March 2020

Qurat-ul-ain Shamsi*
Affiliation:
Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Saeed Ahmad Buzdar
Affiliation:
Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Shagufta Jabeen
Affiliation:
Bioloby Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Khalid Iqbal
Affiliation:
Clinical and Radiation Oncology Department, Shaukat Khanum Memorial Cancer Hospital and Research Centre, Lahore, Pakistan
*
Author for correspondence: Qurat-ul-ain Shamsi, Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan. Tel: +923216827959. E-mail: annieshamsi@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

The emergence of advanced radiotherapy techniques, such as intensity-modulated radiotherapy (IMRT), brachytherapy, conformal radiotherapy, magnetic resonance-guided radiotherapy (MRgRT), stereotactic synchrotron radiotherapy (SSRT) and microbeam radiotherapy (MRT), has increased the importance of the verification of volumetric dose distribution. The verification of dose distribution is usually done by 2D films and 3D gel dosimeters, but PRESAGE® due to its affordability, reproducibility, precision, accuracy, unique dosimetric and physical properties is considered as an effective candidate in providing 3D dose data. PRESAGE® is insensitive to oxygen contamination, machinable and can be molded to a variety of shapes and sizes. It is absorbing rather than scattering light which facilitates high-accuracy readout by optical computed tomography (OP-CT). This review focuses on the feasibility of using PRESAGE® in various complicated radiotherapy techniques by comparing its measured doses with 2D films and treatment planning system (TPS) calculated doses.

Information

Type
Literature Review
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. OD to dose curve for PRESAGE (a); OD to dose curve for EBT2 film (b).

Figure 1

Figure 2. (a) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) and EBT2 film for film 1. (b) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS and EBT2 film for surface film 2.

Figure 2

Figure 3. Gamma maps (a) (±3%/3 mm) between Pinnacle, EBT2 and PRESAGE for a region of PTV 5 mm on film plane; (b) (±3%/3 mm) in the transverse and sagittal planes for PTV 5 mm intersecting film plane.

Figure 3

Figure 4. Line profiles of Pinnacle, PRESAGE and EBT2 film dose distributions from axial slices.

Figure 4

Figure 5. Line profiles of the Pinnacle3, PRESAGE® and EBT2 film dose distributions of axial slice.

Figure 5

Figure 6. GTV, CTV and PTV DVH comparison between the PRESAGE® and Pinnacle3 Planning dose distributions.