Hostname: page-component-74656f9884-npj9w Total loading time: 0 Render date: 2026-11-08T07:48:59.518Z Has data issue: false hasContentIssue false

Analysis of collimator scatter (Sc) in a dual-photon-energy linear accelerator and the effect of scatter factor on extracorporeal irradiation

Published online by Cambridge University Press:  26 August 2026

Giridharan Sampath*
Affiliation:
Department of Radiation Oncology, ClearMedi Radiant Hospital, Mysore, India
Gifty Jebina T
Affiliation:
Department of Medical Physics, Dr. N.G.P. Arts and Science College, Coimbatore, India
Chaitra Katakol
Affiliation:
Department of Radiation Oncology, ClearMedi Radiant Hospital, Mysore, India
Stalin George
Affiliation:
Department of Radiation Oncology, ClearMedi Radiant Hospital, Mysore, India
Daisy G
Affiliation:
Department of Medical Physics, Dr. N.G.P. Arts and Science College, Coimbatore, India
Indhumathi K
Affiliation:
Department of Medical Physics, Dr. N.G.P. Arts and Science College, Coimbatore, India
Sivakumar D
Affiliation:
Department of Medical Physics, Dr. N.G.P. Arts and Science College, Coimbatore, India
*
Corresponding author: Giridharan Sampath; Email: giridharanmedphy@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Aim:

The purpose of this work is to analyse the collimator scatter for photon beams and the effect of scatter factor on extracorporeal irradiation.

Background:

Collimator scatter (Sc) is one of the major components in monitor unit (MU) calculations. A columnar mini-phantom was designed using wax in accordance with the European Society for Radiotherapy and Oncology (ESTRO) recommendations to measure the collimator scatter factor (Sc) for 6 MV and 15 MV photon beams from an Elekta Synergy Platform Linear Accelerator.

Materials and Methods:

Measurements of collimator scatter factors were performed using brass buildup, indigenously designed columnar mini-phantom at depths of dmax and 10 cm to evaluate the influence of electron contamination on Sc. Additionally, measurements were also taken with the columnar mini-phantom at the same depths by placing a 1-mm lead sheet in the accelerator tray. The Sc measured at dmax was higher than that measured at 10 cm depth for both photon energies, indicating the presence of contaminant electrons at shallow depths. The introduction of a 1-mm lead sheet effectively reduced electron contamination during measurements at dmax.

Results:

Wedged fields produced higher Sc values compared with open fields, whereas changes in source-to-surface distance, detector type, chamber orientations and irregular field sizes showed no significant influence on Sc. Measurements obtained using the mini-phantom and a brass buildup cap showed good agreement with each other and were consistent with the AAPM TG-74 recommended data.

Conclusion:

Accurately measured collimator scatter factors and appropriate phantom scatters are essential for reliable MU calculations and accurate dose delivery in extracorporeal irradiation, where phantom scatter is minimal.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2026. Published by Cambridge University Press
Figure 0

Figure 1. Parallel orientation of 6 MV and 15 MV.

Figure 1

Figure 2. Perpendicular orientation of 6 MV and 15 MV.

Figure 2

Figure 3. Brass buildup caps for 6 MV and 15 MV energies used with 0·6 cc and 0·13 cc ionization chambers.

Figure 3

Figure 4. Film calibration curve.

Figure 4

Figure 5. Extracorporeal simulation. Extracorporeal irradiation using Gafchromic film was repeated three times (n = 3) to reduce statistical uncertainity derived from the standard deviation of net optical density was 0·00033.

Figure 5

Figure 6. Variation of Sc with field size at 1·5 cm and 10 cm depth using mini-phantom for 6 MV.

Figure 6

Figure 7. Variation of Sc with field size at 2·7 cm and 10 cm depth using mini-phantom for 15 MV.

Figure 7

Figure 8. Variation of Sc with brass buildup and mini-phantom at 10 cm depth for 6 MV and 15 MV.

Figure 8

Figure 9. Effects of wedges on scatter factor for 6 MV and 15 MV.

Figure 9

Figure 10. Variation of Sc with field sizes at various SSDs for 6 MV and 15 MV.

Figure 10

Figure 11. Variation of Sc for parallel and perpendicular orientations for 6 MV and 15 MV.

Figure 11

Figure 12. Comparison of Sc with two different volume chambers.

Figure 12

Figure 13. Variation of Sc for irregular fields and equivalent fields for 6 MV and 15 MV.

Figure 13

Table 1. Comparison of measured Sc with TG-74 data at 10 cm depthTable 1 long description.

Figure 14

Table 2. Comparison of prescribed and delivered dose measured with Semiflex ionization chamber and Gafchromic filmTable 2 long description.