The Ebola Disinfection Booth: Evaluation of an Enclosed Ultraviolet Light Booth for Disinfection of Contaminated Personal Protective Equipment Prior to Removal

Published online by Cambridge University Press:  20 July 2015

Myreen E. Tomas
Affiliation:
Geriatric Research, Education, and Clinical Center, Cleveland Veterans Affairs Medical Center, Cleveland, Ohio
Jennifer L. Cadnum
Affiliation:
Case Western Reserve University School of Medicine, Cleveland, Ohio
Annette Jencson
Affiliation:
Research Service, Louis Stokes Cleveland Veterans Affairs Medical Center, Cleveland, Ohio
Curtis J. Donskey*
Affiliation:
Geriatric Research, Education, and Clinical Center, Cleveland Veterans Affairs Medical Center, Cleveland, OhioDivision of Infectious Diseases, Department of Medicine, Case Western Reserve University, Cleveland, Ohio.
*
Address correspondence to Curtis J. Donskey, MD, Geriatric Research, Education, and Clinical Center (1110 W), Cleveland Veterans Affairs Medical Center, 10701 East Boulevard, Cleveland, Ohio 44106 (curtisd123@yahoo.com).
Rights & Permissions [Opens in a new window]

Abstract

A portable booth designed to disinfect full-body coverage protective equipment before removal using ultraviolet-C radiation resulted in at least 3 log reductions in bacteriophage MS2 and methicillin-resistant Staphylococcus aureus within 3 minutes. The booth could be useful for disinfection of contaminated protective equipment before removal during care of Ebola patients.

Infect Control Hosp Epidemiol 2015;36(10):1226–1228

Information

Type
Concise Communications
Copyright
© 2015 by The Society for Healthcare Epidemiology of America. All rights reserved 
Figure 0

FIGURE 1The efficacy of an ultraviolet-C radiation disinfection booth for reduction of bacteriophage MS2 (A) and methicillin-resistant Staphylococcus aureus (MRSA) (B) with and (C) without organic load dried onto carriers including glove and gown material and steel disks. The log10 plaque-forming unit (PFU) and log10 colony-forming unit (CFU) reductions per square centimeter are shown for bacteriophage MS2 and MRSA, respectively. Carriers contained 5 to 6 log10 PFU or CFU of the organisms. The carriers were irradiated for 30 seconds or 1, 3, or 5 minutes at a distance of 6 inches from one of the ultraviolet-C lamps or in the center of the booth at 2 feet from each of the 4 ultraviolet-C lamps. The means of the data from experiments conducted in triplicate are presented. Error bars indicate standard error.