Accelerator mass spectrometer facility for heritage radiocarbon dating at NRICH, Korea

Published online by Cambridge University Press:  21 October 2024

Min-Ji Kong
Affiliation:
Conservation Science Division, National Research Institute of Cultural Heritage, 132, Munji-ro, Yuseong-gu, Daejeon, 34122, Republic of Korea
Sujin Park*
Affiliation:
Conservation Science Division, National Research Institute of Cultural Heritage, 132, Munji-ro, Yuseong-gu, Daejeon, 34122, Republic of Korea
Youngeun Kim
Affiliation:
Conservation Science Division, National Research Institute of Cultural Heritage, 132, Munji-ro, Yuseong-gu, Daejeon, 34122, Republic of Korea
*
Corresponding author: Sujin Park; Email: sujin19@korea.kr
Rights & Permissions [Opens in a new window]

Abstract

In 2021, the National Research Institute of Cultural Heritage (NRICH) installed a new accelerator mass spectrometer (AMS) to date Korean national heritage. Before conducting radiocarbon dating, the facility was set up and its performance was assessed. AMS system parameters have been optimized. Measurement of standard and blank samples was verified. Intercomparison analyses were also performed using heritage samples. The F14C value of NIST 4990C was 1.3406 and the background level was 0.0024. Both IAEA-C7 and C8 were confirmed to match the reference value within the 1-σ range. The NIST 4990C measurement results confirmed high precision and accuracy, with 1-σ values below 2‰. In the intercomparison, the error for each sample was 2‰. Thus, based on this study, NRICH plans to launch a dating service specializing in cultural heritage.

Information

Type
Research Article
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona
Figure 0

Figure 1. MICADAS at NRICH.

Figure 1

Table 1. Sample list

Figure 2

Figure 2. Collected wood sample from SMFP Screen in Changdeokgung. (a) Full shot of the sample, including the wood frame with the residual backing paper. (b) and (c) Wood sample used for this study. The red dashed circle in (a) indicates the sampling point. (c) Photograph with a 1 mm scale bar.

Figure 3

Table 2. Two treatment methods, ABA and ABA-B

Figure 4

Table 3. Wood and paper from heritage sample code list

Figure 5

Figure 3. (a) Picture of AGE3 and (b) Diagram of combustion using EA and graphitization using AGE3.

Figure 6

Figure 4. Results of magnet scanning. (a) Results of LE Magnet and (b) HE Magnet. We optimized the condition to center the beam in the beamline. xC indicates stable carbon isotopes, 12C and 13C.

Figure 7

Figure 5. AMS results for NIST 4990C (Ox2). The red dashed line represents the 2‰ error range.

Figure 8

Figure 6. AMS results of background.

Figure 9

Figure 7. AMS results of IAEA standard samples (IAEA-C7&C8). The red dashed line represents the 1-σ error range.

Figure 10

Figure 8. Results of F14C values of IAEA-C5 and C9 for system quality testing. We divide a year into four quarters. They were measured in the 3rd quarter of 2022 and 2023(22Q3–23Q4).

Figure 11

Figure 9. Results of heritage samples as F14C. y-axis is F14C, x-axis is the sample codes. The red solid line represents the mean of data, and the red dotted line represents an error of 2‰.

Figure 12

Table 4. Dating results for heritage samples calibrated by OxCal v4.4.4