Hostname: page-component-74656f9884-j7cmm Total loading time: 0 Render date: 2026-11-07T13:44:32.701Z Has data issue: false hasContentIssue false

Megavoltage versus kilovoltage image guidance for efficiency and accuracy in head and neck IMRT

Published online by Cambridge University Press:  01 December 2009

D. Willis
Affiliation:
1Division of Radiation Oncology
C. Fox
Affiliation:
1Division of Radiation Oncology
A. Haworth
Affiliation:
1Division of Radiation Oncology
A. Rolfo
Affiliation:
1Division of Radiation Oncology
A. Herschtal
Affiliation:
2Biostatistics and Clinical Trials, Peter MacCallum Cancer Centre, St. Andrews Place, Melbourne, Victoria 3002, Australia
T. Kron
Affiliation:
1Division of Radiation Oncology
Rights & Permissions [Opens in a new window]

Abstract

Accurate patient positioning is vitally important in intensity modulated radiation therapy (IMRT) for head and neck (H&N) cancer. The introduction of kilovoltage (kV) on-board imaging (OBI) at our centre was anticipated to improve the accuracy and efficiency of H&N IMRT patient position verification over traditional megavoltage (MV) electronic portal imaging (EPI). This study compares these imaging systems with a phantom accuracy study and retrospective analysis of imaging workload in H&N IMRT at our centre. Six therapists performed online evaluation of phantom images, and residual positional errors for each system were recorded. The largest residual error was 1 mm for OBI and 3 mm for EPI. The estimated improvement in residual error in OBI over EPI was 0.57 mm (95% confidence interval 0.33–0.81 mm), suggesting treatment staff would be better able to detect set-up deviations with OBI. Electronic treatment records of 20 H&N IMRT patients (10 verified daily with MV EPI and 10 with kV OBI) were analysed. Mean imaging session duration was 9.51 min for EPI and 9.76 min for OBI. Analysis found no evidence of an effect on duration due to the imaging system used for this subset of patients (p = 0.664).

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2009
Figure 0

Table 1. Isocentre positions for phantom digitally reconstructed radiograph (DRR) matching sets

Figure 1

Figure 1. Comparison of residual errors in phantom position for kilovoltage on-board imaging (kV OBI) and megavoltage electronic portal imaging (MV EPI) systems. Note that no residual error on kV OBI is larger than 1 mm.

Figure 2

Table 2. The mean and standard deviation of the execution times for head and neck intensity modulated radiation therapy (H&N IMRT) verification imaging on each system

Figure 3

Figure 2. Scatter plot of the proportion of fractions where a correction occurred for each patient against the number of directions where the patient’s individual tolerance for correction was less than 3 mm.